Number of patient samples affected before error detection: Strategic implications for internal quality control and patient-based quality control practices

There is a general impression that patient-based quality control (PBQC) requires a high volume of laboratory results to detect errors effectively. However, internal quality control (IQC) performed infrequently may be associated with increased risk of missed error (i.e. low power of error detection). Using simulations and routine sodium and aspartate aminotransferase (AST) as examples, this study examined how the “average number of patient samples affected before error detection’ (ANPed) metrics can provide linkage to compare relative performance of QC practices in various IQC and PBQC settings. Smaller numbers of IQC samples tested per IQC run or larger average number of patient samples measured between IQC runs are associated with higher ANPed. The ANPed for sodium and AST PBQC models were smaller than IQC performed once every 100 patient samples, except when the systematic error was small for AST. Use of ANPed clearly illustrate the relative impact of different IQC frequencies and number of IQC levels tested. Patient-based quality control can outperform IQC even for laboratories with small testing volume. Laboratory practitioners can use this metric to design a QC strategy that suit their desired risk profile.

Institutions

Publication Details

Journal
Singapore Institute of Technology
Published
2025-01-01
DOI
https://doi.org/10.25447/sit.29587316
Primary Topic
Clinical Laboratory Practices and Quality Control
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Number of patient samples affected before error detection: Strategic implications for internal quality control and patient-based quality control practices

Singapore Institute of Technology
Clinical Laboratory Practices and Quality Control
article

Number of patient samples affected before error detection: Strategic implications for internal quality control and patient-based quality control practices

article en

Abstract

There is a general impression that patient-based quality control (PBQC) requires a high volume of laboratory results to detect errors effectively. However, internal quality control (IQC) performed infrequently may be associated with increased risk of missed error (i.e. low power of error detection). Using simulations and routine sodium and aspartate aminotransferase (AST) as examples, this study examined how the “average number of patient samples affected before error detection’ (ANPed) metrics can provide linkage to compare relative performance of QC practices in various IQC and PBQC settings. Smaller numbers of IQC samples tested per IQC run or larger average number of patient samples measured between IQC runs are associated with higher ANPed. The ANPed for sodium and AST PBQC models were smaller than IQC performed once every 100 patient samples, except when the systematic error was small for AST. Use of ANPed clearly illustrate the relative impact of different IQC frequencies and number of IQC levels tested. Patient-based quality control can outperform IQC even for laboratories with small testing volume. Laboratory practitioners can use this metric to design a QC strategy that suit their desired risk profile.

Singapore Institute of Technology
Singapore Institute of Technology (SG), Flinders University (AU), University Health System (US), National University Health System (SG)
Openalex Percentile: Top 44%
Clinical Laboratory Practices and Quality Control
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.