Establishment of local diagnostic reference levels in terms of dose area product (DAP) in conventional radiology at the Regional Hospital Center of Souss Massa, Morocco

Introduction: The effectiveness of diagnostic reference levels (DRLs) in reducing the doses delivered to patients has been demonstrated. Establishing DRLs in terms of the dose surface product (DSP) is an essential step in the management of radiation doses in conventional radiology. Objective: This retrospective study aimed to assess local diagnostic reference levels (LDRLs) for four types of X-ray examinations in terms of DAP at the Regional Hospital Center of Souss Massa. Materials and methods: Data from 120 adult patients, 30 per location (thorax (PA), pelvis (AP), lumbar spine (AP) and abdomen (AP)) were collected. Patient parameters such as gender, age, BMI, clinical indications, and examination acquisition parameters such as kV, mAs, patient source distance (PSD), and exposure field dimensions were recorded for each patient. The calculation of DRLs is based on a statistical method known as the 75th percentile of the distribution of DAP. The data were statistically analysed by SPSS software V 21.0. Pearson’s parametric test was used to explore the relationship between the different quantitative variables. Results: The mean ESDs for the thorax (PA), lumbar spine (AP), abdomen (AP), and pelvis (AP) were 0.17, 2.55, 2.16, and 2.63 mGy, respectively. The mean DAP values for the thorax (PA), lumbar spine (AP), abdomen (AP), and pelvis (AP) were 16.53, 213.11, 230.55, and 265.19 cGy.cm2, respectively. The DRLs in terms of ESD for the thorax (PA), lumbar spine (AP), abdomen (AP), and pelvis (AP) were 0.17, 2.77, 2.64, 2.87 mGy, respectively. In terms of DAP, they were 16.58, 245, 291.83, and 300.45 cGy.cm2, respectively. The results show a significant relationship between DAP and tension, charge, and ESD for all examinations. In addition, there was a significant relationship between DAP and BMI for abdomen X-ray, and no significant relationship for the three other X-ray examinations. Conclusion: It is possible to reduce the dose delivered to patients in the Regional Hospital Center of Souss Massa through the continuous training of radiology workers, implementation of a quality assurance program for equipment, and institutionalization of DRL as an approach to reducing patient doses and health service costs.

Publication Details

Journal
Springer Link (Chiba Institute of Technology)
Published
2025-09-15
DOI
https://doi.org/10.1051/radiopro/2024062/pdf
Primary Topic
Genetic diversity and population structure
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Establishment of local diagnostic reference levels in terms of dose area product (DAP) in conventional radiology at the Regional Hospital Center of Souss Massa, Morocco

Springer Link (Chiba Institute of Technology)
Genetic diversity and population structure
article

Establishment of local diagnostic reference levels in terms of dose area product (DAP) in conventional radiology at the Regional Hospital Center of Souss Massa, Morocco

article en

Abstract

Introduction: The effectiveness of diagnostic reference levels (DRLs) in reducing the doses delivered to patients has been demonstrated. Establishing DRLs in terms of the dose surface product (DSP) is an essential step in the management of radiation doses in conventional radiology. Objective: This retrospective study aimed to assess local diagnostic reference levels (LDRLs) for four types of X-ray examinations in terms of DAP at the Regional Hospital Center of Souss Massa. Materials and methods: Data from 120 adult patients, 30 per location (thorax (PA), pelvis (AP), lumbar spine (AP) and abdomen (AP)) were collected. Patient parameters such as gender, age, BMI, clinical indications, and examination acquisition parameters such as kV, mAs, patient source distance (PSD), and exposure field dimensions were recorded for each patient. The calculation of DRLs is based on a statistical method known as the 75th percentile of the distribution of DAP. The data were statistically analysed by SPSS software V 21.0. Pearson’s parametric test was used to explore the relationship between the different quantitative variables. Results: The mean ESDs for the thorax (PA), lumbar spine (AP), abdomen (AP), and pelvis (AP) were 0.17, 2.55, 2.16, and 2.63 mGy, respectively. The mean DAP values for the thorax (PA), lumbar spine (AP), abdomen (AP), and pelvis (AP) were 16.53, 213.11, 230.55, and 265.19 cGy.cm2, respectively. The DRLs in terms of ESD for the thorax (PA), lumbar spine (AP), abdomen (AP), and pelvis (AP) were 0.17, 2.77, 2.64, 2.87 mGy, respectively. In terms of DAP, they were 16.58, 245, 291.83, and 300.45 cGy.cm2, respectively. The results show a significant relationship between DAP and tension, charge, and ESD for all examinations. In addition, there was a significant relationship between DAP and BMI for abdomen X-ray, and no significant relationship for the three other X-ray examinations. Conclusion: It is possible to reduce the dose delivered to patients in the Regional Hospital Center of Souss Massa through the continuous training of radiology workers, implementation of a quality assurance program for equipment, and institutionalization of DRL as an approach to reducing patient doses and health service costs.

Springer Link (Chiba Institute of Technology)
Openalex Percentile: Top 45%
Genetic diversity and population structure
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Establishment of local diagnostic reference levels in terms of dose area product (DAP) in conventional radiology at the Regional Hospital Center of Souss Massa, Morocco · Springer Link (Chiba Institute of Technology) (2025) | TGRS Research Map | TGRS