Feasibility of an Ultra-Low-Dose PET Scan Protocol with CT-Based and LSO-TX-Based Attenuation Correction Using a Long-Axial-Field-of-View PET/CT Scanner.

Long-axial-field-of-view (LAFOV) PET scanners enable substantial reduction in injected radiotracer activity while maintaining clinically feasible scan times. Whole-body CT scans performed for PET attenuation correction can significantly add to total radiation exposure. We investigated the feasibility of an ultra-low-dose PET protocol and the application of a CT-less PET attenuation correction method (lutetium oxyorthosilicate background transmission [LSO-TX]) that uses 176Lu background radiation from detector scintillators with low-count PET data. Methods: Each of the 4 study subjects was scanned for 90 min using an ultra-low-dose 18F-FDG protocol (injected activity, 6.7-9.0 MBq) with an LAFOV PET scanner. PET images were reconstructed with different frame durations using low-dose CT-based and LSO-TX-based attenuation maps (μ-maps). The image quality of PET images was assessed by the signal-to-noise ratio (SNR) in the liver and the contrast-to-noise ratio in the brain. Absolute errors in SUVs between PET images reconstructed with LSO-TX-based and CT-based μ-maps were assessed at each scan duration. Results: Visual assessment showed that 20-30 min of PET data obtained using 18F-FDG activities below 10 MBq (i.e., 0.1 MBq/kg) can yield high-quality images. PET images reconstructed with CT-based and LSO-TX-based μ-maps had comparable SNRs and contrast-to-noise ratios at all scan durations. The mean ± SD SNRs of PET images reconstructed with the CT-based and the LSO-TX-based μ-maps were 9.2 ± 1.9 dB and 9.8 ± 2.0 dB at 90-min scan duration, 6.8 ± 1.7 dB and 6.9 ± 1.8 dB at 30-min scan duration, and 5.5 ± 1.2 dB and 5.6 ± 1.2 dB at 20-min scan duration, respectively. The relative absolute SUV errors between PET images reconstructed with LSO-TX-based and CT-based μ-maps ranged from 3.1% to 6.4% across different volumes of interest with a 20-min scan duration. Conclusion: PET scans with an LAFOV scanner maintained good visual image quality with 18F-FDG activities below 10 MBq for scan durations of 20-30 min. The LSO-TX-based attenuation correction method yielded images comparable to those obtained with the CT-based attenuation correction method in such protocols.

Publication Details

Journal
Open Access CRIS of the University of Bern
Published
2025-05-05
DOI
https://doi.org/10.48620/87787
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Feasibility of an Ultra-Low-Dose PET Scan Protocol with CT-Based and LSO-TX-Based Attenuation Correction Using a Long-Axial-Field-of-View PET/CT Scanner.

Open Access CRIS of the University of Bern
Medical Imaging Techniques and Applications
article

Feasibility of an Ultra-Low-Dose PET Scan Protocol with CT-Based and LSO-TX-Based Attenuation Correction Using a Long-Axial-Field-of-View PET/CT Scanner.

article en

Abstract

Long-axial-field-of-view (LAFOV) PET scanners enable substantial reduction in injected radiotracer activity while maintaining clinically feasible scan times. Whole-body CT scans performed for PET attenuation correction can significantly add to total radiation exposure. We investigated the feasibility of an ultra-low-dose PET protocol and the application of a CT-less PET attenuation correction method (lutetium oxyorthosilicate background transmission [LSO-TX]) that uses 176Lu background radiation from detector scintillators with low-count PET data. Methods: Each of the 4 study subjects was scanned for 90 min using an ultra-low-dose 18F-FDG protocol (injected activity, 6.7-9.0 MBq) with an LAFOV PET scanner. PET images were reconstructed with different frame durations using low-dose CT-based and LSO-TX-based attenuation maps (μ-maps). The image quality of PET images was assessed by the signal-to-noise ratio (SNR) in the liver and the contrast-to-noise ratio in the brain. Absolute errors in SUVs between PET images reconstructed with LSO-TX-based and CT-based μ-maps were assessed at each scan duration. Results: Visual assessment showed that 20-30 min of PET data obtained using 18F-FDG activities below 10 MBq (i.e., 0.1 MBq/kg) can yield high-quality images. PET images reconstructed with CT-based and LSO-TX-based μ-maps had comparable SNRs and contrast-to-noise ratios at all scan durations. The mean ± SD SNRs of PET images reconstructed with the CT-based and the LSO-TX-based μ-maps were 9.2 ± 1.9 dB and 9.8 ± 2.0 dB at 90-min scan duration, 6.8 ± 1.7 dB and 6.9 ± 1.8 dB at 30-min scan duration, and 5.5 ± 1.2 dB and 5.6 ± 1.2 dB at 20-min scan duration, respectively. The relative absolute SUV errors between PET images reconstructed with LSO-TX-based and CT-based μ-maps ranged from 3.1% to 6.4% across different volumes of interest with a 20-min scan duration. Conclusion: PET scans with an LAFOV scanner maintained good visual image quality with 18F-FDG activities below 10 MBq for scan durations of 20-30 min. The LSO-TX-based attenuation correction method yielded images comparable to those obtained with the CT-based attenuation correction method in such protocols.

Open Access CRIS of the University of Bern
Openalex Percentile: Top 39%
Medical Imaging Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Feasibility of an Ultra-Low-Dose PET Scan Protocol with CT-Based and LSO-TX-Based Attenuation Correction Using a Long-Axial-Field-of-View PET/CT Scanner. · Open Access CRIS of the University of Bern (2025) | TGRS Research Map | TGRS