Analysis and consideration of radiation dose distribution in cardiac catheterization interventions

Objective: To assess the variations in radiation dose at different areas in the operation room of cardiac catheterization laboratory during cardiac catheterization interventions. Methods: To simulate the conventional operation in cardiac catheterization interventions, perform angiography on standard manikins, the radiation dose was collected from 22 areas in the operation room of cardiac catheterization laboratory under 8 projection angles, and each area was repeated 5 times, and the collected data was analyzed by one-way ANOVA and independent sample t-test. Results: Analysis of the radiation dose in 22 areas under 8 projection angles revealed that the lowest radiation dose was found in the area at the end of the operating table in the cardiac catheterization room (p < 0.05), and the highest radiation dose was found in the area on the left and right sides of the X-ray tube (p < 0.05); the radiation dose in the area on the right side of the X-ray tube of the DSA machine was greater than that in the conventional standing area of the first operator (p < 0.05), and the radiation dose in the conventional standing area of the second operator was greater than that in the conventional standing area of the first operator (p < 0.05); the radiation dose in the standing area of the first operator was highest in the cephalic position (CRA 30°) (p < 0.05), and the radiation dose in the standing area of the second operator was highest in the left anterior oblique position (LAO 45°) (p < 0.05); in the angle of RAO projection, the radiation dose in the right area of the X-ray tube was greater than that in the left area of the X-ray tube (p < 0.05), and the result was reversed at the projection angle of the LAO. Conclusion: The radiation dose during cardiac catheterization interventions is lowest in the area at the end of the operating table, which can be used as a standing area for nurses and as an area for the placement of surgical equipment and supplies. At the same time, it is necessary to pay attention to the radiation dose to the second operator and to further improve the radiation protection measures for the second operator, and additional measures are needed to minimize the radiation dose to the operators in the cephalic position (CRA 30°) and in the left anterior oblique position (LAO 45°).

Publication Details

Journal
Springer Link (Chiba Institute of Technology)
Published
2025-03-14
DOI
https://doi.org/10.1051/radiopro/2024037/pdf
Primary Topic
Mycorrhizal Fungi and Plant Interactions
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Analysis and consideration of radiation dose distribution in cardiac catheterization interventions

Springer Link (Chiba Institute of Technology)
Mycorrhizal Fungi and Plant Interactions
article

Analysis and consideration of radiation dose distribution in cardiac catheterization interventions

article en

Abstract

Objective: To assess the variations in radiation dose at different areas in the operation room of cardiac catheterization laboratory during cardiac catheterization interventions. Methods: To simulate the conventional operation in cardiac catheterization interventions, perform angiography on standard manikins, the radiation dose was collected from 22 areas in the operation room of cardiac catheterization laboratory under 8 projection angles, and each area was repeated 5 times, and the collected data was analyzed by one-way ANOVA and independent sample t-test. Results: Analysis of the radiation dose in 22 areas under 8 projection angles revealed that the lowest radiation dose was found in the area at the end of the operating table in the cardiac catheterization room (p < 0.05), and the highest radiation dose was found in the area on the left and right sides of the X-ray tube (p < 0.05); the radiation dose in the area on the right side of the X-ray tube of the DSA machine was greater than that in the conventional standing area of the first operator (p < 0.05), and the radiation dose in the conventional standing area of the second operator was greater than that in the conventional standing area of the first operator (p < 0.05); the radiation dose in the standing area of the first operator was highest in the cephalic position (CRA 30°) (p < 0.05), and the radiation dose in the standing area of the second operator was highest in the left anterior oblique position (LAO 45°) (p < 0.05); in the angle of RAO projection, the radiation dose in the right area of the X-ray tube was greater than that in the left area of the X-ray tube (p < 0.05), and the result was reversed at the projection angle of the LAO. Conclusion: The radiation dose during cardiac catheterization interventions is lowest in the area at the end of the operating table, which can be used as a standing area for nurses and as an area for the placement of surgical equipment and supplies. At the same time, it is necessary to pay attention to the radiation dose to the second operator and to further improve the radiation protection measures for the second operator, and additional measures are needed to minimize the radiation dose to the operators in the cephalic position (CRA 30°) and in the left anterior oblique position (LAO 45°).

Springer Link (Chiba Institute of Technology)
Good health and well-being
Openalex Percentile: Top 53%
Mycorrhizal Fungi and Plant Interactions
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Analysis and consideration of radiation dose distribution in cardiac catheterization interventions · Springer Link (Chiba Institute of Technology) (2025) | TGRS Research Map | TGRS