Comparison of water-equivalent and effective diameter-based size-specific dose estimates in computed tomography pulmonary angiography: implications for radiation dose optimization

This retrospective study investigates dose estimation in computed tomography pulmonary angiography (CTPA) by analyzing 200 examinations from 96 male and 104 female patients. Two size-specific dose estimation (SSDE) methods were evaluated: one based on effective diameter (SSDEDeff) and the other on water-equivalent diameter (SSDEDw). The effective diameter (Deff) was manually measured using the lateral (LAT) and anteroposterior (AP) dimensions, while the water-equivalent diameter (Dw) was calculated from the region of interest (ROI) using mean Hounsfield Unit (HU) values and cross-sectional area. SSDE calculations were compared to the conventional CT dose metric, CTDIvol, to assess its underestimation relative to SSDEDw. The findings revealed a dose underestimation of up to 18% for CTDIvol compared to SSDEDw, particularly for patients with larger body habitus. SSDEDeff consistently overestimated the dose by 7.73% compared to SSDEDw, as Deff relied only on external dimensions while Dw considered tissue attenuation for a more individualized dose assessment. High correlations were observed between lateral diameter (dLAT) and both Deff (R2 = 0.9175) and Dw (R2 = 0.7578). However, the systematic overestimation by SSDEDeff emphasizes the importance of clearly specifying the metric used for SSDE, as differences can influence reported doses by 5–10%, affecting clinical dose monitoring and adherence to diagnostic reference levels. This study highlights the limitations of CTDIvol and SSDEDeff compared to SSDEDw, especially in regions with heterogeneous tissue. The results support the use of individualized radiation dose assessments to improve radiation safety, particularly in CTPA, which has seen increased application during the COVID-19 pandemic.

Publication Details

Journal
Springer Link (Chiba Institute of Technology)
Published
2025-09-15
DOI
https://doi.org/10.1051/radiopro/2024063/pdf
Primary Topic
Mycorrhizal Fungi and Plant Interactions
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparison of water-equivalent and effective diameter-based size-specific dose estimates in computed tomography pulmonary angiography: implications for radiation dose optimization

Springer Link (Chiba Institute of Technology)
Mycorrhizal Fungi and Plant Interactions
article

Comparison of water-equivalent and effective diameter-based size-specific dose estimates in computed tomography pulmonary angiography: implications for radiation dose optimization

article en

Abstract

This retrospective study investigates dose estimation in computed tomography pulmonary angiography (CTPA) by analyzing 200 examinations from 96 male and 104 female patients. Two size-specific dose estimation (SSDE) methods were evaluated: one based on effective diameter (SSDEDeff) and the other on water-equivalent diameter (SSDEDw). The effective diameter (Deff) was manually measured using the lateral (LAT) and anteroposterior (AP) dimensions, while the water-equivalent diameter (Dw) was calculated from the region of interest (ROI) using mean Hounsfield Unit (HU) values and cross-sectional area. SSDE calculations were compared to the conventional CT dose metric, CTDIvol, to assess its underestimation relative to SSDEDw. The findings revealed a dose underestimation of up to 18% for CTDIvol compared to SSDEDw, particularly for patients with larger body habitus. SSDEDeff consistently overestimated the dose by 7.73% compared to SSDEDw, as Deff relied only on external dimensions while Dw considered tissue attenuation for a more individualized dose assessment. High correlations were observed between lateral diameter (dLAT) and both Deff (R2 = 0.9175) and Dw (R2 = 0.7578). However, the systematic overestimation by SSDEDeff emphasizes the importance of clearly specifying the metric used for SSDE, as differences can influence reported doses by 5–10%, affecting clinical dose monitoring and adherence to diagnostic reference levels. This study highlights the limitations of CTDIvol and SSDEDeff compared to SSDEDw, especially in regions with heterogeneous tissue. The results support the use of individualized radiation dose assessments to improve radiation safety, particularly in CTPA, which has seen increased application during the COVID-19 pandemic.

Springer Link (Chiba Institute of Technology)
Good health and well-being
Openalex Percentile: Top 54%
Mycorrhizal Fungi and Plant Interactions
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.