Biomechanical Comparison of Titanium and Stainless Steel Miniplates / Screws for Sagittal Split Osteotomy- An in Vitro Study

Background: Studies on titanium and stainless-steel miniplates in Oral and Maxillofacial Surgery have been published, but none have compared their biomechanical properties for sagittal split osteotomy fixation. The objectives of the present study were to investigate the biochemical efficacy of titanium mini-plates and to examine the effectiveness of titanium mini-plates in comparison with stainless steel plates in the management of sagittal split osteotomy fixation. Methods: The study involved 10 dry cadaveric hemimandibles, grouped into two groups: Group-SS (stainless) and Group-TT (titanium). Each mandible underwent sagittal split osteotomy and was treated with stainless steel and titanium miniplates. Biochemical testing was performed using the Instron Corporation Series IX automated system. Data were analyzed using SPSS version 16. Results: The two-tailed probability-test showed a significant difference in parameters like "load at break", "displacement at maximum load", and "compressive strength" between 10 hemi-mandible specimens compared to stainless steel. The study revealed that stainless steel miniplates had higher maximum load values and displacement compared to titanium miniplates, with varying compressive strength and displacement. The mean maximum load was 501.20N ± 106.01 and displacement was 13.98 mm ± 3.83 mm. Conclusions: Stainless-steel miniplates were found to be ideal for the management of sagittal split osteotomy. Stainless steel miniplates provided satisfactory osteosynthesis and were cost effective compared to titanium plates. Titanium plates are more malleable and can be easily adapted to the varying contours of the facial skeleton which clinically translates into the reduced time required for plating.

Institutions

Publication Details

Journal
Zenodo (CERN European Organization for Nuclear Research)
Published
2025-02-28
DOI
https://doi.org/10.5281/zenodo.15202249
Primary Topic
Spatial and Panel Data Analysis
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Biomechanical Comparison of Titanium and Stainless Steel Miniplates / Screws for Sagittal Split Osteotomy- An in Vitro Study

Zenodo (CERN European Organization for Nuclear Research)
Spatial and Panel Data Analysis
article

Biomechanical Comparison of Titanium and Stainless Steel Miniplates / Screws for Sagittal Split Osteotomy- An in Vitro Study

article en

Abstract

Background: Studies on titanium and stainless-steel miniplates in Oral and Maxillofacial Surgery have been published, but none have compared their biomechanical properties for sagittal split osteotomy fixation. The objectives of the present study were to investigate the biochemical efficacy of titanium mini-plates and to examine the effectiveness of titanium mini-plates in comparison with stainless steel plates in the management of sagittal split osteotomy fixation. Methods: The study involved 10 dry cadaveric hemimandibles, grouped into two groups: Group-SS (stainless) and Group-TT (titanium). Each mandible underwent sagittal split osteotomy and was treated with stainless steel and titanium miniplates. Biochemical testing was performed using the Instron Corporation Series IX automated system. Data were analyzed using SPSS version 16. Results: The two-tailed probability-test showed a significant difference in parameters like "load at break", "displacement at maximum load", and "compressive strength" between 10 hemi-mandible specimens compared to stainless steel. The study revealed that stainless steel miniplates had higher maximum load values and displacement compared to titanium miniplates, with varying compressive strength and displacement. The mean maximum load was 501.20N ± 106.01 and displacement was 13.98 mm ± 3.83 mm. Conclusions: Stainless-steel miniplates were found to be ideal for the management of sagittal split osteotomy. Stainless steel miniplates provided satisfactory osteosynthesis and were cost effective compared to titanium plates. Titanium plates are more malleable and can be easily adapted to the varying contours of the facial skeleton which clinically translates into the reduced time required for plating.

Zenodo (CERN European Organization for Nuclear Research)
International Association for Dental Research (US)
Openalex Percentile: Top 45%
Spatial and Panel Data Analysis
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.