RESEARCH OF THE INFLUENCE OF MECHANICAL INFLUENCES ON THERMOCATALYTIC SENSOR OPERATION
To experimentally evaluate the possible consequences for gas analyzers of impact loads that do not lead to failure of the catalytic thermometer. It has been revealed that with significant impacts on thermocatalytic sensors that do not lead to sensor failure, a significant shift in the zero of the measuring bridge of gas analyzers occurs while maintaining their constant sensitivity. In this case, the shift in zero can be both in the direction of positive and negative values.
Institutions
- Dnipro University of Technology (UA)
Publication Details
- Journal
- Zenodo (CERN European Organization for Nuclear Research)
- Published
- 2024-12-29
- DOI
- https://doi.org/10.5281/zenodo.14569037
- Primary Topic
- Material Science and Thermodynamics
- Type
- article
- Field-Weighted Citation Impact
- 0.00