Determination of Scatter Radiation Round Three Different Models of Mammography Unit

Abstract : Purpose: this study was carried out to determine the amount of scatter radiation around three different models of digital mammography units may contribute to shielding calculations. Objectives: measuring and comparing for scattered radiation at four orientations (4 x 90o=360o) round three mammography models. Materials and methods: RADOS (RDS-120 Universal Survey Meter) device was used to measure scatter radiation at four orientations round three types of mammography digital models NM-GA, Liyum, and Mammomat. These orientations are backward, forward, left lateral, and right lateral. All measurements are taken in Craniocaudal projection. Results: backward orientation has highest and right lateral has lowest intensity of scatter radiations when three models of digital mammography were used. Mammomat model with backward orientation was at highest 6.5μSv of intensity and NM-GA model at lowest intensity of 1.1 μSv from right lateral orientation. Conclusions: The obtained measurements of scatter radiation at four different orientations may be used in a shielding calculation when NM-GA, Liyum, and Mammomat of mammography models are used.

Institutions

Publication Details

Journal
Zenodo (CERN European Organization for Nuclear Research)
Published
2025-02-05
DOI
https://doi.org/10.5281/zenodo.14812023
Primary Topic
Digital Radiography and Breast Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Determination of Scatter Radiation Round Three Different Models of Mammography Unit

Zenodo (CERN European Organization for Nuclear Research)
Digital Radiography and Breast Imaging
article

Determination of Scatter Radiation Round Three Different Models of Mammography Unit

article en

Abstract

Abstract : Purpose: this study was carried out to determine the amount of scatter radiation around three different models of digital mammography units may contribute to shielding calculations. Objectives: measuring and comparing for scattered radiation at four orientations (4 x 90o=360o) round three mammography models. Materials and methods: RADOS (RDS-120 Universal Survey Meter) device was used to measure scatter radiation at four orientations round three types of mammography digital models NM-GA, Liyum, and Mammomat. These orientations are backward, forward, left lateral, and right lateral. All measurements are taken in Craniocaudal projection. Results: backward orientation has highest and right lateral has lowest intensity of scatter radiations when three models of digital mammography were used. Mammomat model with backward orientation was at highest 6.5μSv of intensity and NM-GA model at lowest intensity of 1.1 μSv from right lateral orientation. Conclusions: The obtained measurements of scatter radiation at four different orientations may be used in a shielding calculation when NM-GA, Liyum, and Mammomat of mammography models are used.

Zenodo (CERN European Organization for Nuclear Research)
Al-Hussein Bin Talal University (JO), University of Medical Sciences and Technology (SD), Sudan University of Science and Technology (SD)
Openalex Percentile: Top 38%
Digital Radiography and Breast Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.