VIEshunt: towards a ventricular intelligent and electromechanical shunt for hydrocephalus therapy

BACKGROUND Shunt systems for hydrocephalus therapy are commonly based on passive mechanical pressure valves that are driven by the intracranial, intra-abdominal, and hydrostatic pressure. The differential pressure acting on the valve determines the drainage rate of cerebrospinal fluid (CSF) but is not a gauge of the physiological condition of the patient. Internal and external influences can cause over- or underdrainage and lead to pathological levels of intracranial pressure (ICP). METHODS The first prototype of a ventricular intelligent and electromechanical shunt (VIEshunt) is developed, tested, and compared with previous efforts towards the development of a smart shunt. Its key components are a micro pump, a flow meter, a pressure sensor, an inertial measurement unit, a wireless communication interface, and a microcontroller. The VIEshunt prototype was tested in vitro using a hardware-in-the-loop (HiL) test bench that runs real-time patient simulations involving changes in intracranial and intra-abdominal pressure, as well as changes in posture ranging between supine and upright position. The prototype was subsequently tested in an in vivo pilot study based on an acute ovine animal model (n=1) with infusions of artificial CSF. RESULTS During 24 h in vitro testing, the prototype detected the simulated posture changes of the patient and automatically adapted the controller reference. The posture-specific ICP references of 12 mmHg for supine and -3 mmHg for upright position were tracked without offset, thus preventing adverse over- and underdrainage during the investigated HiL test scenario. During acute in vivo testing, the prototype first regulated the mean ICP of a sheep from 22 mmHg down to 20 mmHg. Each of the three subsequent intraventricular bolus infusions of 1 mL saline solution increased mean ICP by approximately 11 mmHg. While natural absorption alone decreased ICP by only 5 mmHg within 9 min, the prototype was able to regulate ICP back to the pre-bolus pressure value within 5 min. CONCLUSION The developed VIEshunt prototype is capable of posture-dependent ICP regulation and CSF drainage control. Smart shunt systems based on VIEshunt could improve patient monitoring and enable optimal physiologic therapy by adapting to the individual patient. To derive statistically relevant conclusions for the performance of VIEshunt, future work will focus on the development of a next generation prototype for use in pre-clinical studies.

Publication Details

Journal
Zurich Open Repository and Archive (University of Zurich)
Published
2025-03-14
DOI
https://doi.org/10.5167/uzh-276513
Primary Topic
Molecular Biology Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

VIEshunt: towards a ventricular intelligent and electromechanical shunt for hydrocephalus therapy

Zurich Open Repository and Archive (University of Zurich)
Molecular Biology Techniques and Applications
article

VIEshunt: towards a ventricular intelligent and electromechanical shunt for hydrocephalus therapy

article en

Abstract

BACKGROUND Shunt systems for hydrocephalus therapy are commonly based on passive mechanical pressure valves that are driven by the intracranial, intra-abdominal, and hydrostatic pressure. The differential pressure acting on the valve determines the drainage rate of cerebrospinal fluid (CSF) but is not a gauge of the physiological condition of the patient. Internal and external influences can cause over- or underdrainage and lead to pathological levels of intracranial pressure (ICP). METHODS The first prototype of a ventricular intelligent and electromechanical shunt (VIEshunt) is developed, tested, and compared with previous efforts towards the development of a smart shunt. Its key components are a micro pump, a flow meter, a pressure sensor, an inertial measurement unit, a wireless communication interface, and a microcontroller. The VIEshunt prototype was tested in vitro using a hardware-in-the-loop (HiL) test bench that runs real-time patient simulations involving changes in intracranial and intra-abdominal pressure, as well as changes in posture ranging between supine and upright position. The prototype was subsequently tested in an in vivo pilot study based on an acute ovine animal model (n=1) with infusions of artificial CSF. RESULTS During 24 h in vitro testing, the prototype detected the simulated posture changes of the patient and automatically adapted the controller reference. The posture-specific ICP references of 12 mmHg for supine and -3 mmHg for upright position were tracked without offset, thus preventing adverse over- and underdrainage during the investigated HiL test scenario. During acute in vivo testing, the prototype first regulated the mean ICP of a sheep from 22 mmHg down to 20 mmHg. Each of the three subsequent intraventricular bolus infusions of 1 mL saline solution increased mean ICP by approximately 11 mmHg. While natural absorption alone decreased ICP by only 5 mmHg within 9 min, the prototype was able to regulate ICP back to the pre-bolus pressure value within 5 min. CONCLUSION The developed VIEshunt prototype is capable of posture-dependent ICP regulation and CSF drainage control. Smart shunt systems based on VIEshunt could improve patient monitoring and enable optimal physiologic therapy by adapting to the individual patient. To derive statistically relevant conclusions for the performance of VIEshunt, future work will focus on the development of a next generation prototype for use in pre-clinical studies.

Zurich Open Repository and Archive (University of Zurich)
Openalex Percentile: Top 58%
Molecular Biology Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.