Effects of shortening distal shunt catheters in children as relates to patient age and valve type: a simulation study with clinical implications.

Objective In children treated with a ventriculoperitoneal shunt (VPS), the growing body length increases the negative hydrostatic pressure within the shunt. At initial shunt insertion in neonates, infants, and small children, some surgeons traditionally shorten the distal catheter, although full length is generally well tolerated. Authors of this study investigate the effect of shortening distal shunt catheters on intracranial pressure (ICP) and cerebrospinal fluid (CSF) drainage rates in a bench test model set for different ages and with particular valve types.Methods Using a hardware-in-the-loop test bench, three different VPS settings were evaluated with both a standard 120-cm-long distal catheter and a short distal catheter (SDC) of 60 cm: a miniNav differential pressure valve with a 10-cmH2O opening pressure (MN10) and adjustable proGAV valves with a 25-cmH2O gravitational unit set at 5-cmH2O (PG5) and 10-cmH2O (PG10) opening pressures. The hardware was adjusted for the body length of 1-, 5-, and 10-year-old children. The software simulated the age-adapted physiological conditions of intracranial and intra-abdominal pressures. All valve and catheter combinations were tested 5 times, each in 1-hour runs in three positions: first horizontal, vertical, and second horizontal. Flow through the VPS and intracranial pressure were measured.Results The use of an SDC, as compared to the standard catheter, led to a significant ICP decrease in the vertical position using the PG5 across all age groups, PG10 for 5 year olds, and MN10 for 10 year olds. Using the SDC also resulted in a significantly higher drained CSF volume in the vertical position across all age groups with the MN10 and PG5.Conclusions Shortening the distal catheter leads to increased drainage rates with significant ICP decreases in a 1-year-old child model setup. This effect is more pronounced for low opening pressure valves, unprotected valve systems, and model setups for older children. These results suggest that an SDC may promote overdrainage of variable magnitude depending on the age and height of the child and the valve setting. Therefore, an SDC should be avoided.

Institutions

Publication Details

Journal
Open Access CRIS of the University of Bern
Published
2025-07-15
DOI
https://doi.org/10.48620/89500
Primary Topic
Microscopic Colitis
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Effects of shortening distal shunt catheters in children as relates to patient age and valve type: a simulation study with clinical implications.

Open Access CRIS of the University of Bern
Microscopic Colitis
article

Effects of shortening distal shunt catheters in children as relates to patient age and valve type: a simulation study with clinical implications.

article en

Abstract

Objective In children treated with a ventriculoperitoneal shunt (VPS), the growing body length increases the negative hydrostatic pressure within the shunt. At initial shunt insertion in neonates, infants, and small children, some surgeons traditionally shorten the distal catheter, although full length is generally well tolerated. Authors of this study investigate the effect of shortening distal shunt catheters on intracranial pressure (ICP) and cerebrospinal fluid (CSF) drainage rates in a bench test model set for different ages and with particular valve types.Methods Using a hardware-in-the-loop test bench, three different VPS settings were evaluated with both a standard 120-cm-long distal catheter and a short distal catheter (SDC) of 60 cm: a miniNav differential pressure valve with a 10-cmH2O opening pressure (MN10) and adjustable proGAV valves with a 25-cmH2O gravitational unit set at 5-cmH2O (PG5) and 10-cmH2O (PG10) opening pressures. The hardware was adjusted for the body length of 1-, 5-, and 10-year-old children. The software simulated the age-adapted physiological conditions of intracranial and intra-abdominal pressures. All valve and catheter combinations were tested 5 times, each in 1-hour runs in three positions: first horizontal, vertical, and second horizontal. Flow through the VPS and intracranial pressure were measured.Results The use of an SDC, as compared to the standard catheter, led to a significant ICP decrease in the vertical position using the PG5 across all age groups, PG10 for 5 year olds, and MN10 for 10 year olds. Using the SDC also resulted in a significantly higher drained CSF volume in the vertical position across all age groups with the MN10 and PG5.Conclusions Shortening the distal catheter leads to increased drainage rates with significant ICP decreases in a 1-year-old child model setup. This effect is more pronounced for low opening pressure valves, unprotected valve systems, and model setups for older children. These results suggest that an SDC may promote overdrainage of variable magnitude depending on the age and height of the child and the valve setting. Therefore, an SDC should be avoided.

Open Access CRIS of the University of Bern
Klinik und Poliklinik für Psychiatrie und Psychotherapie (DE)
Openalex Percentile: Top 44%
Microscopic Colitis
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Effects of shortening distal shunt catheters in children as relates to patient age and valve type: a simulation study with clinical implications. · Open Access CRIS of the University of Bern (2025) | TGRS Research Map | TGRS