Influence of scan extension and starting quadrant on the accuracy of four intraoral scanners for fabricating tooth-supported crowns.

STATEMENT OF PROBLEM Multiple factors can influence the accuracy of intraoral scanners (IOSs). However, the impact of scan extension and starting quadrant on the accuracy of IOSs for fabricating tooth-supported crowns remains uncertain. PURPOSE The purpose of the present in vitro study was to measure the influence of scan extension (half or complete arch scan) and the starting quadrant (same quadrant or contralateral quadrant of the location of the crown preparation) on the accuracy of four IOSs. MATERIAL AND METHODS A typodont with a crown preparation on the left first molar was digitized (T710) to obtain a reference scan. Four scanner groups were created: TRIOS 5, PrimeScan, i700, and iTero. Then, 3 subgroups were defined based on the scan extension and starting quadrant: half arch (HA subgroup), complete arch scan starting on the left quadrant (CA-same subgroup), and complete arch scan starting on the right quadrant (CA-contralateral subgroup), (n=15). The reference scan was used as a control to measure the root mean square (RMS) error discrepancies with each experimental scan on the tooth preparation, margin of the tooth preparation, and adjacent tooth areas. Two-way ANOVA and pairwise multiple comparisons were used to analyze trueness (α=.05). The Levene and pairwise comparisons using the Wilcoxon Rank sum tests were used to analyze precision (α=.05). RESULTS For the tooth preparation analysis, significant trueness and precision differences were found among the groups (P<.001) and subgroups (P<.001), with a significant interaction group×subgroup (P=.002). The iTero and TRIOS5 groups obtained better trueness than the PrimeScan and i700 groups (P<.001). Moreover, half arch scans obtained the best trueness, while the CA-contralateral scans obtained the worst trueness (P<.001). The iTero group showed the worst precision among the IOSs tested. For the margin of the tooth preparation evaluation, significant trueness and precision differences were found among the groups (P<.001) and subgroups (P<.001), with a significant interaction group×subgroup (P=.005). The iTero group obtained best trueness (P<.001), but the worst precision (P<.001) among the IOSs tested. Half arch scans obtained the best trueness and precision values. For the adjacent tooth analysis, trueness and precision differences were found among the groups (P<.001) and subgroups tested (P<.001), with a significant interaction group×subgroup (P=.005). The TRIOS 5 obtained the best trueness and precision. Half arch scans obtained the best accuracy. CONCLUSIONS Scan extension and the starting quadrant impacted the scanning trueness and precision of the IOSs tested. Additionally, the IOSs showed varying scanning discrepancies depending on the scanning area assessed. Half arch scans presented the highest trueness and precision, and the complete arch scans in which the scan started in the contralateral quadrant of where the crown preparation was obtained the worst trueness and precision.

Institutions

Publication Details

Journal
Open Access CRIS of the University of Bern
Published
2024-10-26
DOI
https://doi.org/10.48350/196128
Primary Topic
Dental materials and restorations
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Influence of scan extension and starting quadrant on the accuracy of four intraoral scanners for fabricating tooth-supported crowns.

Open Access CRIS of the University of Bern
Dental materials and restorations
article

Influence of scan extension and starting quadrant on the accuracy of four intraoral scanners for fabricating tooth-supported crowns.

article en

Abstract

STATEMENT OF PROBLEM Multiple factors can influence the accuracy of intraoral scanners (IOSs). However, the impact of scan extension and starting quadrant on the accuracy of IOSs for fabricating tooth-supported crowns remains uncertain. PURPOSE The purpose of the present in vitro study was to measure the influence of scan extension (half or complete arch scan) and the starting quadrant (same quadrant or contralateral quadrant of the location of the crown preparation) on the accuracy of four IOSs. MATERIAL AND METHODS A typodont with a crown preparation on the left first molar was digitized (T710) to obtain a reference scan. Four scanner groups were created: TRIOS 5, PrimeScan, i700, and iTero. Then, 3 subgroups were defined based on the scan extension and starting quadrant: half arch (HA subgroup), complete arch scan starting on the left quadrant (CA-same subgroup), and complete arch scan starting on the right quadrant (CA-contralateral subgroup), (n=15). The reference scan was used as a control to measure the root mean square (RMS) error discrepancies with each experimental scan on the tooth preparation, margin of the tooth preparation, and adjacent tooth areas. Two-way ANOVA and pairwise multiple comparisons were used to analyze trueness (α=.05). The Levene and pairwise comparisons using the Wilcoxon Rank sum tests were used to analyze precision (α=.05). RESULTS For the tooth preparation analysis, significant trueness and precision differences were found among the groups (P<.001) and subgroups (P<.001), with a significant interaction group×subgroup (P=.002). The iTero and TRIOS5 groups obtained better trueness than the PrimeScan and i700 groups (P<.001). Moreover, half arch scans obtained the best trueness, while the CA-contralateral scans obtained the worst trueness (P<.001). The iTero group showed the worst precision among the IOSs tested. For the margin of the tooth preparation evaluation, significant trueness and precision differences were found among the groups (P<.001) and subgroups (P<.001), with a significant interaction group×subgroup (P=.005). The iTero group obtained best trueness (P<.001), but the worst precision (P<.001) among the IOSs tested. Half arch scans obtained the best trueness and precision values. For the adjacent tooth analysis, trueness and precision differences were found among the groups (P<.001) and subgroups tested (P<.001), with a significant interaction group×subgroup (P=.005). The TRIOS 5 obtained the best trueness and precision. Half arch scans obtained the best accuracy. CONCLUSIONS Scan extension and the starting quadrant impacted the scanning trueness and precision of the IOSs tested. Additionally, the IOSs showed varying scanning discrepancies depending on the scanning area assessed. Half arch scans presented the highest trueness and precision, and the complete arch scans in which the scan started in the contralateral quadrant of where the crown preparation was obtained the worst trueness and precision.

Open Access CRIS of the University of Bern
University of Dental Medicine (MM)
Openalex Percentile: Top 61%
Dental materials and restorations
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.