Power-Efficient Control of Non-Linear Magnetic Field Generators for MPI
The scaling of electrical power constitutes a significant challenge when adapting Magnetic Particle Imaging (MPI) to a human scale. The use of coils incorporating soft-iron cores serves to reduce power usage, but also introduces spatial imperfections and non-linearities in the current-to-field relationship. This study proposes methodologies for the control of the magnetic field output of a system comprising 18 coils, subject to the influence of saturated iron. In particular, we integrate current sequence optimization with neural network-based predictions for field and gradient values, thereby enabling the precise and power-optimal generation of magnetic fields. The proposed framework for controlling non-linear magnetic field generators represents a significant advancement in MPI technology, paving the way for the development of human-scale, power-efficient medical imaging solutions.
Institutions
- Universität Hamburg (DE)
- MSH Medical School Hamburg – University of Applied Sciences and Medical University (DE)
- University Medical Center Hamburg-Eppendorf (DE)
Publication Details
- Journal
- Fraunhofer-Publica (Fraunhofer-Gesellschaft)
- Published
- 2024-11-12
- DOI
- https://doi.org/10.18416/ijmpi.2025.2503023
- Primary Topic
- Animal and Plant Science Education
- Type
- article
- Field-Weighted Citation Impact
- 0.00