CapacitanceMethod for Eliminating Gas IonizationNoise in X‑ray Detection Performance Testing
The exploration of semiconductive materials for direct X-ray detection has aroused intense interest recently, and many perovskites and MOFs claimed to have excellent X-ray detection abilities. However, a recent report indicates that gas ionization introduces significant interference when evaluating the X-ray detection performance under ambient conditions by the widely used two-probe method. This work reveals that the use of a capacitance method may eliminate the gas ionization noise of semiconductor detectors by about 98% (noise signal ratio: CdTe, ∼0; CsPbBr<sub>3</sub>, ∼2%; MAPbBr<sub>3</sub>, ∼2%). This method does not require high vacuum or encapsulation and is highly suggested for the valuation of X-ray direct detection performance.
Publication Details
- Journal
- Figshare
- Published
- 2025-03-07
- DOI
- https://doi.org/10.1021/acsmaterialslett.4c02567.s001
- Primary Topic
- Art, Technology, and Culture
- Type
- article
- Field-Weighted Citation Impact
- 0.00