Non-invasive follow-up for intracranial aneurysms treated with contour neurovascular system-comparison of digital subtraction angiography (DSA) to magnetic resonance imaging (MRI) and spectral computed tomography angiography (CTA) in vitro

PURPOSE The contour neurovascular system (CNS) is an intrasaccular flow-disrupting device designed for the treatment of intracranial wide-necked bifurcation aneurysms. Metal artifacts limit magnetic resonance imaging (MRI) accessibility after implantation. The purpose of this in vitro study was to evaluate non-invasive imaging alternatives to digital subtraction angiography (DSA). MATERIAL AND METHODS Three aneurysms of patients originally treated with CNS were three-dimensional (3D)-printed (one at the basilar tip and two at the middle cerebral artery bifurcation). CNS devices were implanted under fluoroscopic control into the 3D models. Post-implantation two-dimensional-DSA, flat panel computed tomography angiography (CTA), MRI, and spectral CTA were performed. RESULTS Time of flight angiography and T1 weighted sequences showed large susceptibility artifacts at the detachment zone of the devices. A thin-sliced T2 weighted sequence in cross-sectional orientation to the aneurysm allowed visualization of the aneurysm dome, but the aneurysm neck and parent vessel could not be assessed. Focused spectral CTA, especially a 40 keV reconstruction with a metal artifact reduction algorithm (orthopedic metal artifact reduction (OMAR)), showed only minor artifacts at the detachment zone. This approach achieved a very similar result to DSA and flat panel computed tomography, enabling the assessment of the device structure, aneurysm perfusion, and parent vessel perfusion. DISCUSSION AND CONCLUSION For non-invasive follow-up of CNS, focused 40 keV CTA with OMAR seems to be a valuable option. MRI can be valuable for larger aneurysms to assess the aneurysm dome, but was not suitable for evaluating the parent vessels and aneurysm neck after CNS implantation in this study.

Publication Details

Journal
Zurich Open Repository and Archive (University of Zurich)
Published
2024-09-02
DOI
https://doi.org/10.5167/uzh-268484
Primary Topic
Intracranial Aneurysms: Treatment and Complications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Non-invasive follow-up for intracranial aneurysms treated with contour neurovascular system-comparison of digital subtraction angiography (DSA) to magnetic resonance imaging (MRI) and spectral computed tomography angiography (CTA) in vitro

Zurich Open Repository and Archive (University of Zurich)
Intracranial Aneurysms: Treatment and Complications
article

Non-invasive follow-up for intracranial aneurysms treated with contour neurovascular system-comparison of digital subtraction angiography (DSA) to magnetic resonance imaging (MRI) and spectral computed tomography angiography (CTA) in vitro

article en

Abstract

PURPOSE The contour neurovascular system (CNS) is an intrasaccular flow-disrupting device designed for the treatment of intracranial wide-necked bifurcation aneurysms. Metal artifacts limit magnetic resonance imaging (MRI) accessibility after implantation. The purpose of this in vitro study was to evaluate non-invasive imaging alternatives to digital subtraction angiography (DSA). MATERIAL AND METHODS Three aneurysms of patients originally treated with CNS were three-dimensional (3D)-printed (one at the basilar tip and two at the middle cerebral artery bifurcation). CNS devices were implanted under fluoroscopic control into the 3D models. Post-implantation two-dimensional-DSA, flat panel computed tomography angiography (CTA), MRI, and spectral CTA were performed. RESULTS Time of flight angiography and T1 weighted sequences showed large susceptibility artifacts at the detachment zone of the devices. A thin-sliced T2 weighted sequence in cross-sectional orientation to the aneurysm allowed visualization of the aneurysm dome, but the aneurysm neck and parent vessel could not be assessed. Focused spectral CTA, especially a 40 keV reconstruction with a metal artifact reduction algorithm (orthopedic metal artifact reduction (OMAR)), showed only minor artifacts at the detachment zone. This approach achieved a very similar result to DSA and flat panel computed tomography, enabling the assessment of the device structure, aneurysm perfusion, and parent vessel perfusion. DISCUSSION AND CONCLUSION For non-invasive follow-up of CNS, focused 40 keV CTA with OMAR seems to be a valuable option. MRI can be valuable for larger aneurysms to assess the aneurysm dome, but was not suitable for evaluating the parent vessels and aneurysm neck after CNS implantation in this study.

Zurich Open Repository and Archive (University of Zurich)
Openalex Percentile: Top 55%
Intracranial Aneurysms: Treatment and Complications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Non-invasive follow-up for intracranial aneurysms treated with contour neurovascular system-comparison of digital subtraction angiography (DSA) to magnetic resonance imaging (MRI) and spectral computed tomography angiography (CTA) in vitro · Zurich Open Repository and Archive (University of Zurich) (2024) | TGRS Research Map | TGRS