Influence of Normal Tissue Objective Tools on Treatment Planning System in Nasopharyngeal Carcinoma (NPC): A 3D Printed Anthropomorphic Phantom Study

Introduction: The normal tissue objective (NTO), one of the new aspects in the radiation treatment planning system (TPS), aims to lower the absorbed dose received by organs at risk (OARs) close to the target volume or Planning Target Volume (PTV). This study was conducted to ascertain the impact of planning in nasopharyngeal cancer (NPC) cases both with and without manual NTO settings.Material and Methods: The study used a 3D printed head and neck phantom exposed to radiation using the Intensity Modulated Radiation Therapy (IMRT) technique with 6000 cGy prescribed dose and divided into 30 fractions to find the discrepancies between the manually calculated absorbed dose­ and the automatic calculated absorbed dose of TPS. Moreover, evaluation parameter indicators, including the homogeneity index (HI), conformity index (CI), gradient index (GI), and comprehensive quality index (CQI), were used to make comparisons. NTO parameter used in manual plans are f0 = 107%, f∞ = 65%, dose fall-off = 0.05 mm-1, and xstart = 0.75 cm.Results: The statistical analysis resulted in a significant difference between the calculated absorbed dose and TPS's absorbed dose of Automatic NTO and Manual NTO, whereas, Without NTO plans, there was no statistical difference. The HI values for Automatic NTO, Manual NTO, and Without NTO are 0.118, 0.05 , and 0.053, respectively. The CI values for Automatic NTO, Manual NTO, and Without NTO are 0.91, 0.99, and 0,19. The GI value for Automatic NTO, Manual NTO, and Without NTO are 3.34, 4.94, and 7.29, respectively. CQI parameter showed that the Automatic NTO plan performs better than the Manual NTO plan based on the maximum dose received by the OAR.Conclusion: In this study, the manual NTO plan showed better performance by reducing hot spots in the central region of PTV.

Publication Details

Journal
SHILAP Revista de lepidopterología
Published
2024-12-01
DOI
https://doi.org/10.22038/ijmp.2024.76906.2352
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Influence of Normal Tissue Objective Tools on Treatment Planning System in Nasopharyngeal Carcinoma (NPC): A 3D Printed Anthropomorphic Phantom Study

SHILAP Revista de lepidopterología
Advanced Radiotherapy Techniques
article

Influence of Normal Tissue Objective Tools on Treatment Planning System in Nasopharyngeal Carcinoma (NPC): A 3D Printed Anthropomorphic Phantom Study

article en

Abstract

Introduction: The normal tissue objective (NTO), one of the new aspects in the radiation treatment planning system (TPS), aims to lower the absorbed dose received by organs at risk (OARs) close to the target volume or Planning Target Volume (PTV). This study was conducted to ascertain the impact of planning in nasopharyngeal cancer (NPC) cases both with and without manual NTO settings.Material and Methods: The study used a 3D printed head and neck phantom exposed to radiation using the Intensity Modulated Radiation Therapy (IMRT) technique with 6000 cGy prescribed dose and divided into 30 fractions to find the discrepancies between the manually calculated absorbed dose­ and the automatic calculated absorbed dose of TPS. Moreover, evaluation parameter indicators, including the homogeneity index (HI), conformity index (CI), gradient index (GI), and comprehensive quality index (CQI), were used to make comparisons. NTO parameter used in manual plans are f0 = 107%, f∞ = 65%, dose fall-off = 0.05 mm-1, and xstart = 0.75 cm.Results: The statistical analysis resulted in a significant difference between the calculated absorbed dose and TPS's absorbed dose of Automatic NTO and Manual NTO, whereas, Without NTO plans, there was no statistical difference. The HI values for Automatic NTO, Manual NTO, and Without NTO are 0.118, 0.05 , and 0.053, respectively. The CI values for Automatic NTO, Manual NTO, and Without NTO are 0.91, 0.99, and 0,19. The GI value for Automatic NTO, Manual NTO, and Without NTO are 3.34, 4.94, and 7.29, respectively. CQI parameter showed that the Automatic NTO plan performs better than the Manual NTO plan based on the maximum dose received by the OAR.Conclusion: In this study, the manual NTO plan showed better performance by reducing hot spots in the central region of PTV.

SHILAP Revista de lepidopterología
Good health and well-being
Openalex Percentile: Top 61%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.