Delta-P as an early indicator of oxygenator failure: the case for standardized manufacturer reference values

The evolution of extracorporeal membrane oxygenation (ECMO) has led to an increasing reliance on objective parameters to detect complications early and enhance patient safety. One such parameter is the transmembrane pressure gradient, more commonly referred to as delta P (ΔP), the difference between the inlet and outlet pressures of the oxygenator. This measure has increasingly been recognized as a critical early indicator of thrombus formation within the oxygenator during ECMO support. However, its interpretation remains complex and context-dependent, particularly because delta P values are significantly influenced by the design features of different ECMO circuits, including pump head geometry and oxygenator configuration.

Publication Details

Journal
Springer Link (Chiba Institute of Technology)
Published
2025-12-17
DOI
https://doi.org/10.1051/ject/2025039/pdf
Primary Topic
Mechanical Circulatory Support Devices
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Delta-P as an early indicator of oxygenator failure: the case for standardized manufacturer reference values

Springer Link (Chiba Institute of Technology)
Mechanical Circulatory Support Devices
article

Delta-P as an early indicator of oxygenator failure: the case for standardized manufacturer reference values

article en

Abstract

The evolution of extracorporeal membrane oxygenation (ECMO) has led to an increasing reliance on objective parameters to detect complications early and enhance patient safety. One such parameter is the transmembrane pressure gradient, more commonly referred to as delta P (ΔP), the difference between the inlet and outlet pressures of the oxygenator. This measure has increasingly been recognized as a critical early indicator of thrombus formation within the oxygenator during ECMO support. However, its interpretation remains complex and context-dependent, particularly because delta P values are significantly influenced by the design features of different ECMO circuits, including pump head geometry and oxygenator configuration.

Springer Link (Chiba Institute of Technology)
Openalex Percentile: Top 58%
Mechanical Circulatory Support Devices
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.