EVALUATION OF X-RAY EQUIPMENT PERFORMANCE IN SOME GOVERNMENT HOSPITALS IN KADUNA CITY, KADUNA, NIGERIA

Background: In low- and middle-income countries, imaging equipment is frequently left unmonitored for extended periods of time, and only during inspections for licensing purposes is its quality assessed in terms of performance evaluation and quality control Objective: The objective of this study was to conduct an evaluation of X-ray equipment performance in some government hospitals in Kaduna city, Kaduna state, Nigeria Method: In this investigation, three (3) facilities were deliberately visited. At each hospital, Xray equipment was subjected to quality control (QC) testing, and the X-ray rooms were physically inspected. Hospitals execute the following quality control tests: radiation dosage levels, radiation leakage, half value layer, beam alignment, peak tube voltage, mA linearity, and entrance skin dose. Using a survey meter (RADOS model RDS-120), the radiation dose levels in each X-ray room were measured. Results: The acquired data revealed that only Hospital 1 and Hospital 2 exhibited a high level of compliance. Hospital 1 and Hospital 2 passed the radiation leakage test with results below the tolerance limit of 1.0 mSv/hr, demonstrating compliance with the norm. Hospital 3 passed the Half-value layer test, and all three hospitals passed the beam alignment test. Hospitals 1 and 2 passed the peak tube voltage (kVp) test because the co-efficient of variation between the set kVp and measured kVp was less than 0.05. Only Hospital 2 passed the mA linearity test, since its coefficient of linearity was 0.08, which is below the permitted level of 0.1. The outcome of the entrance skin dose (ESD) test revealed that all of the hospitals fell under the international and national tolerance limits, which is deemed safe for both patients and employees. Conclusion: Based on the findings of this study, it can be concluded that the government hospitals in the Kaduna city are slightly in compliance with national and international standards. The visual inspection performed on the hospitals revealed that they retain a high degree of compliance to the standard.

Institutions

Publication Details

Journal
JRRS
Published
2025-05-16
DOI
https://doi.org/10.82547/jrrs.vol37no2.18
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

EVALUATION OF X-RAY EQUIPMENT PERFORMANCE IN SOME GOVERNMENT HOSPITALS IN KADUNA CITY, KADUNA, NIGERIA

JRRS
Radiation Dose and Imaging
article

EVALUATION OF X-RAY EQUIPMENT PERFORMANCE IN SOME GOVERNMENT HOSPITALS IN KADUNA CITY, KADUNA, NIGERIA

article en

Abstract

Background: In low- and middle-income countries, imaging equipment is frequently left unmonitored for extended periods of time, and only during inspections for licensing purposes is its quality assessed in terms of performance evaluation and quality control Objective: The objective of this study was to conduct an evaluation of X-ray equipment performance in some government hospitals in Kaduna city, Kaduna state, Nigeria Method: In this investigation, three (3) facilities were deliberately visited. At each hospital, Xray equipment was subjected to quality control (QC) testing, and the X-ray rooms were physically inspected. Hospitals execute the following quality control tests: radiation dosage levels, radiation leakage, half value layer, beam alignment, peak tube voltage, mA linearity, and entrance skin dose. Using a survey meter (RADOS model RDS-120), the radiation dose levels in each X-ray room were measured. Results: The acquired data revealed that only Hospital 1 and Hospital 2 exhibited a high level of compliance. Hospital 1 and Hospital 2 passed the radiation leakage test with results below the tolerance limit of 1.0 mSv/hr, demonstrating compliance with the norm. Hospital 3 passed the Half-value layer test, and all three hospitals passed the beam alignment test. Hospitals 1 and 2 passed the peak tube voltage (kVp) test because the co-efficient of variation between the set kVp and measured kVp was less than 0.05. Only Hospital 2 passed the mA linearity test, since its coefficient of linearity was 0.08, which is below the permitted level of 0.1. The outcome of the entrance skin dose (ESD) test revealed that all of the hospitals fell under the international and national tolerance limits, which is deemed safe for both patients and employees. Conclusion: Based on the findings of this study, it can be concluded that the government hospitals in the Kaduna city are slightly in compliance with national and international standards. The visual inspection performed on the hospitals revealed that they retain a high degree of compliance to the standard.

JRRS
Ahmadu Bello University (NG)
Openalex Percentile: Top 42%
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.