Efficacy of Air Powder Water-Jet Devices in Cleaning Implant Surfaces in a Non-Surgical Peri-Implantitis Treatment Simulation-A Laboratory Study.

Objective To evaluate the impact of implant surface and instrumentation time on the efficacy of two air powder water-jet (APWJ) devices in cleaning the implant surface in a simulation of non-surgical peri-implantitis treatment.Materials And Methods Turned and modified surface implants (28 each) were coated with a biofilm imitation and mounted on resin models replicating peri-implant intra-osseous defects, including a soft tissue replica. The entire implant periphery was instrumented for 5 or 15s per implant sextant (i.e., in total 30 or 90s per implant), with one of two different APWJ devices using either a glycine or an erythritol powder. Residual biofilm imitation was automatically assessed on standardized photographs and expressed as percentage of the exposed implant surface.Results Implant surface (ε2:0.253, p < 0.001) and instrumentation time (ε2:0.044, p = 0.036) had a moderate and small effect, respectively, on the outcome, that is, instrumenting turned compared to modified surface implants as well as using a longer compared to a shorter instrumentation time resulted in less residual biofilm imitation. Complete biofilm imitation removal was achieved only in four turned implants, treated for 15s per sextant. Every second turned implant presented with a maximum of 5% residual biofilm imitation, while only two modified implants achieved this level of cleanliness.Conclusion In a non-surgical peri-implantitis treatment simulation with APWJ devices, superior biofilm imitation removal was achieved at turned implants, and a longer instrumentation time resulted in less residual biofilm imitation. Modified implants had high chances of incomplete biofilm imitation removal, especially at the apical part of the defect. Complete biofilm imitation removal was in general largely unpredictable.

Institutions

Publication Details

Journal
Open Access CRIS of the University of Bern
Published
2025-06-20
DOI
https://doi.org/10.48620/88635
Primary Topic
Erosion and Abrasive Machining
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Efficacy of Air Powder Water-Jet Devices in Cleaning Implant Surfaces in a Non-Surgical Peri-Implantitis Treatment Simulation-A Laboratory Study.

Open Access CRIS of the University of Bern
Erosion and Abrasive Machining
article

Efficacy of Air Powder Water-Jet Devices in Cleaning Implant Surfaces in a Non-Surgical Peri-Implantitis Treatment Simulation-A Laboratory Study.

article en

Abstract

Objective To evaluate the impact of implant surface and instrumentation time on the efficacy of two air powder water-jet (APWJ) devices in cleaning the implant surface in a simulation of non-surgical peri-implantitis treatment.Materials And Methods Turned and modified surface implants (28 each) were coated with a biofilm imitation and mounted on resin models replicating peri-implant intra-osseous defects, including a soft tissue replica. The entire implant periphery was instrumented for 5 or 15s per implant sextant (i.e., in total 30 or 90s per implant), with one of two different APWJ devices using either a glycine or an erythritol powder. Residual biofilm imitation was automatically assessed on standardized photographs and expressed as percentage of the exposed implant surface.Results Implant surface (ε2:0.253, p < 0.001) and instrumentation time (ε2:0.044, p = 0.036) had a moderate and small effect, respectively, on the outcome, that is, instrumenting turned compared to modified surface implants as well as using a longer compared to a shorter instrumentation time resulted in less residual biofilm imitation. Complete biofilm imitation removal was achieved only in four turned implants, treated for 15s per sextant. Every second turned implant presented with a maximum of 5% residual biofilm imitation, while only two modified implants achieved this level of cleanliness.Conclusion In a non-surgical peri-implantitis treatment simulation with APWJ devices, superior biofilm imitation removal was achieved at turned implants, and a longer instrumentation time resulted in less residual biofilm imitation. Modified implants had high chances of incomplete biofilm imitation removal, especially at the apical part of the defect. Complete biofilm imitation removal was in general largely unpredictable.

Open Access CRIS of the University of Bern
University of Dental Medicine (MM)
Clean water and sanitation
Openalex Percentile: Top 77%
Erosion and Abrasive Machining
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.