Investigating the performance of polytetrafluoroethylene solid free radical dosimeter
This study leverages the ability of polytetrafluoroethylene (PTFE) to generate stable free radicals under ionizing radiation to develop a solid PTFE dosimeter sheet. Experimental results indicate a strong positive correlation between free radical response and absorbed dose over a broad range (1‒180 kGy). The performance of the dosimeter was minimally influenced by dose rate and energy dependence, while the effects of irradiation temperature and post-irradiation storage time on free-radical decay exhibited clear and consistent patterns. Unlike conventional dosimeters used in the industry, the PTFE dosimeter is suitable for high-dose measurements and calibrations, covering a range of 1‒180 kGy with a total uncertainty of less than 5.02%, demonstrating high metrological reliability. This study provides valuable insights into precise dose control and calibration in the irradiation processing of polymers and other functional materials.
Institutions
- Chinese Academy of Sciences (CN)
- Shanghai Institute of Applied Physics (CN)
- Shanghai Civil Aviation College (CN)
Publication Details
- Journal
- SHILAP Revista de lepidopterología
- Published
- 2024-12-01
- DOI
- https://doi.org/10.11889/j.1000-3436.2024-0106
- Citations
- 1
- Primary Topic
- Radiation Effects and Dosimetry
- Type
- article
- Field-Weighted Citation Impact
- 0.35