Free-air ionization chambers for the measurement of air kerma in low-energy x-rays—optimum air path length and the limitations of averaging monoenergetic correction factors

Abstract The German national metrology institute Physikalisch-Technische Bundesanstalt is developing a new primary standard for the realization of the quantity air kerma free-in-air in the x-ray energy range up to 50 keV. National metrology institutes use free-air ionization chambers (FACs) for this purpose. FACs with various geometrical dimensions are used for low-energy x-rays. This study aims to (1) determine the optimum air path length of a FAC, which is the most critical geometric parameter for the low-energy x-ray range, and (2) to evaluate the limitations of averaging monoenergetic correction factors, a method that many institutes use to enable time-saving calculation of spectral correction factors for a larger number of radiation qualities. Monte Carlo simulations were performed to compute monoenergetic and spectral correction factors for air attenuation, photon scattering, inadequate charged particle equilibrium, and diaphragm effects. In addition, spectral correction factors were calculated by averaging the monoenergetic correction factors with different x-ray spectra. The results show that for low-energy FACs the predominant correction is the correction for air attenuation. Thus, it is preferable to have a minimized air path length. The minimization of the air path length is limited by the impact of electrons created by diaphragm interactions and the loss of charged particle equilibrium. If the air attenuation between the aperture and the collecting volume is considered for a 10 kV x-ray spectrum, the averaged and simulated spectral correction factors agree within less than 0.1%.

Publication Details

Journal
Metrologia
Published
2025-03-21
DOI
https://doi.org/10.1088/1681-7575/adc39d
Citations
1
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
1.88
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Free-air ionization chambers for the measurement of air kerma in low-energy x-rays—optimum air path length and the limitations of averaging monoenergetic correction factors

1 citations
Metrologia
Advanced Radiotherapy Techniques
1.88
article

Free-air ionization chambers for the measurement of air kerma in low-energy x-rays—optimum air path length and the limitations of averaging monoenergetic correction factors

article en
1 citations

Abstract

Abstract The German national metrology institute Physikalisch-Technische Bundesanstalt is developing a new primary standard for the realization of the quantity air kerma free-in-air in the x-ray energy range up to 50 keV. National metrology institutes use free-air ionization chambers (FACs) for this purpose. FACs with various geometrical dimensions are used for low-energy x-rays. This study aims to (1) determine the optimum air path length of a FAC, which is the most critical geometric parameter for the low-energy x-ray range, and (2) to evaluate the limitations of averaging monoenergetic correction factors, a method that many institutes use to enable time-saving calculation of spectral correction factors for a larger number of radiation qualities. Monte Carlo simulations were performed to compute monoenergetic and spectral correction factors for air attenuation, photon scattering, inadequate charged particle equilibrium, and diaphragm effects. In addition, spectral correction factors were calculated by averaging the monoenergetic correction factors with different x-ray spectra. The results show that for low-energy FACs the predominant correction is the correction for air attenuation. Thus, it is preferable to have a minimized air path length. The minimization of the air path length is limited by the impact of electrons created by diaphragm interactions and the loss of charged particle equilibrium. If the air attenuation between the aperture and the collecting volume is considered for a 10 kV x-ray spectrum, the averaged and simulated spectral correction factors agree within less than 0.1%.

MetrologiaVol. 62(2)
Affordable and clean energy
Openalex Percentile: Top 11%
Advanced Radiotherapy Techniques
1.88
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.