Diagnostic Value of Ultrasound Elastography Combined with BI-RADS in the Risk Stratification of Breast Masses

Background: The incorporation of elastography into standard ultrasound assessments increases the confidence of radiologist in opting for short-term monitoring instead of unnecessary biopsies for questionable breast lesions. Aim: This study aimed to assess the diagnostic performance of ultrasound BI-RADS combined with elastography parameters [elastography score (ES) and strain ratio (SR)] for differentiating benign from malignant breast lesions. Patients and methods: A retrospective review was conducted on 137 solid breast lesions classified as BI-RADS 3 or 4 on ultrasound. Histopathology served as the reference standard. Diagnostic indices including accuracy, sensitivity, specificity, positive and negative predictive values, and the area under the ROC curve (AUC) were calculated for BI-RADS alone and in combination with elastography measures. Results: Of the examined lesions, 77 were benign and 60 were malignant. Patients with malignant tumors were significantly older (p < 0.001). Malignant masses demonstrated higher ES and SR values and were more frequently assigned to higher BI-RADS categories (p < 0.001). Adding elastography to BI-RADS improved sensitivity from 89.9% to 100%, though specificity declined slightly from 92% to 89%. Compared to pairing BI-RADS with ES, the combination of BI-RADS with SR produced higher sensitivity (95% vs. 90%) but lower specificity (86% vs. 92%). Both combined approaches yielded higher PPV than BI-RADS alone (90% and 92% vs. 89%). Using SR cutoff of 2.45 produced 89.9% sensitivity and 69.2% specificity, while ES cutoff of 3.5 yielded 45.8% sensitivity and 80% specificity. Conclusion: Combining BI-RADS with ES or SR substantially enhanced the detection of malignant breast lesions. This integrated approach improved diagnostic confidence and helped limit unnecessary biopsies by refining risk stratification.

Publication Details

Journal
The Egyptian Journal of Hospital Medicine
Published
2025-10-01
DOI
https://doi.org/10.21608/ejhm.2025.471413
Primary Topic
Ultrasound Imaging and Elastography
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Diagnostic Value of Ultrasound Elastography Combined with BI-RADS in the Risk Stratification of Breast Masses

The Egyptian Journal of Hospital Medicine
Ultrasound Imaging and Elastography
article

Diagnostic Value of Ultrasound Elastography Combined with BI-RADS in the Risk Stratification of Breast Masses

article en

Abstract

Background: The incorporation of elastography into standard ultrasound assessments increases the confidence of radiologist in opting for short-term monitoring instead of unnecessary biopsies for questionable breast lesions. Aim: This study aimed to assess the diagnostic performance of ultrasound BI-RADS combined with elastography parameters [elastography score (ES) and strain ratio (SR)] for differentiating benign from malignant breast lesions. Patients and methods: A retrospective review was conducted on 137 solid breast lesions classified as BI-RADS 3 or 4 on ultrasound. Histopathology served as the reference standard. Diagnostic indices including accuracy, sensitivity, specificity, positive and negative predictive values, and the area under the ROC curve (AUC) were calculated for BI-RADS alone and in combination with elastography measures. Results: Of the examined lesions, 77 were benign and 60 were malignant. Patients with malignant tumors were significantly older (p < 0.001). Malignant masses demonstrated higher ES and SR values and were more frequently assigned to higher BI-RADS categories (p < 0.001). Adding elastography to BI-RADS improved sensitivity from 89.9% to 100%, though specificity declined slightly from 92% to 89%. Compared to pairing BI-RADS with ES, the combination of BI-RADS with SR produced higher sensitivity (95% vs. 90%) but lower specificity (86% vs. 92%). Both combined approaches yielded higher PPV than BI-RADS alone (90% and 92% vs. 89%). Using SR cutoff of 2.45 produced 89.9% sensitivity and 69.2% specificity, while ES cutoff of 3.5 yielded 45.8% sensitivity and 80% specificity. Conclusion: Combining BI-RADS with ES or SR substantially enhanced the detection of malignant breast lesions. This integrated approach improved diagnostic confidence and helped limit unnecessary biopsies by refining risk stratification.

The Egyptian Journal of Hospital MedicineVol. 101(1)
Openalex Percentile: Top 43%
Ultrasound Imaging and Elastography
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.