Fluoroscopy Exposure During Distal Radius Fracture Fixation: An Audit of Current Practice

Background Distal radius fractures (DRFs) are among the most common fractures in adults. Volar plate fixation is a widely used treatment. Intraoperative fluoroscopy is essential for confirming reduction and implant positioning, but it exposes patients and staff to ionising radiation. This Study was conducted to assess local practice, quantify radiation exposure, and help to establish national benchmarks for fluoroscopy usage and radiation exposure to improve patients' and staff safety. Methods A retrospective review of 106 consecutive patients undergoing volar plate fixation for DRFs was performed in William Harvey Hospital, Ashford, Kent, UK. Demographics, fracture type, surgical details, and intraoperative fluoroscopy parameters were collected, including screening time, number of exposures, dose area product (DAP), and air kerma (AK). Data were analysed by fracture type and surgeon grade. Results The cohort included 31 males and 75 females, with a mean age of 59 years. Most fractures were intra-articular (n = 93). All patients underwent volar plate fixation. The mean operative time was 73 minutes. Fluoroscopy exposure during distal radius fixation demonstrated a mean screening time of 24 seconds (range 0.1-111), with a mean of 43 exposures (range 5-209), a mean dose area product (DAP) of 4.75 µGy·m² (range 0.31-29.26), and a mean AK of 0.26 (range 0.01-9). Further subgroup analysis was performed to quantify exposure by surgeons' grade and fracture classification. Conclusion Intraoperative fluoroscopy during volar plate fixation of DRFs in our institution resulted in a mean DAP of 4.75 µGy·m² and a mean screening time of 24 seconds. No significant differences were observed by operator grade or fracture complexity. Future efforts should focus on standardising its collection and defining orthopaedic Diagnostic reference levels. Regular audit of fluoroscopy usage against established standards would help in improving uniformity across hospitals and orthopaedic teams.

Authors

Institutions

Publication Details

Journal
Cureus
Published
2025-12-20
DOI
https://doi.org/10.7759/cureus.99719
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Fluoroscopy Exposure During Distal Radius Fracture Fixation: An Audit of Current Practice

Nabil Seoudi, Mohammad Hassan Murad, Mutmainah Ekungba-Adewole, Siddharth Virani et al.
Cureus
Radiation Dose and Imaging
article

Fluoroscopy Exposure During Distal Radius Fracture Fixation: An Audit of Current Practice

Nabil Seoudi, Mohammad Hassan Murad, Mutmainah Ekungba-Adewole, Siddharth Virani, Tarik Al-Dahan, Salih Seidahmed, Yogan Gurung, Aliye Nazli Asardag, Idahosa Izedonmwen, Nima Sherpa
article en

Abstract

Background Distal radius fractures (DRFs) are among the most common fractures in adults. Volar plate fixation is a widely used treatment. Intraoperative fluoroscopy is essential for confirming reduction and implant positioning, but it exposes patients and staff to ionising radiation. This Study was conducted to assess local practice, quantify radiation exposure, and help to establish national benchmarks for fluoroscopy usage and radiation exposure to improve patients' and staff safety. Methods A retrospective review of 106 consecutive patients undergoing volar plate fixation for DRFs was performed in William Harvey Hospital, Ashford, Kent, UK. Demographics, fracture type, surgical details, and intraoperative fluoroscopy parameters were collected, including screening time, number of exposures, dose area product (DAP), and air kerma (AK). Data were analysed by fracture type and surgeon grade. Results The cohort included 31 males and 75 females, with a mean age of 59 years. Most fractures were intra-articular (n = 93). All patients underwent volar plate fixation. The mean operative time was 73 minutes. Fluoroscopy exposure during distal radius fixation demonstrated a mean screening time of 24 seconds (range 0.1-111), with a mean of 43 exposures (range 5-209), a mean dose area product (DAP) of 4.75 µGy·m² (range 0.31-29.26), and a mean AK of 0.26 (range 0.01-9). Further subgroup analysis was performed to quantify exposure by surgeons' grade and fracture classification. Conclusion Intraoperative fluoroscopy during volar plate fixation of DRFs in our institution resulted in a mean DAP of 4.75 µGy·m² and a mean screening time of 24 seconds. No significant differences were observed by operator grade or fracture complexity. Future efforts should focus on standardising its collection and defining orthopaedic Diagnostic reference levels. Regular audit of fluoroscopy usage against established standards would help in improving uniformity across hospitals and orthopaedic teams.

CureusVol. 17(12)
Maidstone and Tunbridge Wells NHS Trust (GB), William Harvey Hospital (GB), East Kent Hospitals University NHS Foundation Trust (GB), Tunbridge Wells Hospital (GB)
Openalex Percentile: Top 43%
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.