Accessing Higher Order Mathieu Space Stability Zones to Narrow Isolation Widths Using Digital Quadrupole Time-of-Flight Mass Spectrometry

Digital quadrupoles are driven with rectangular RF waveforms and, through duty cycle control, can access higher-order Mathieu space stability zones (HZs) using comparatively low voltages. Analytically, accessing these zones remains attractive as HZs exhibit higher resolving powers ((<i>m</i>/<i>z</i>)/(Δ<i>m</i>/<i>z</i>)) compared to the conventional quadrupole operation in zone (1,1). Presented here is a modification of a commercial quadrupole time-of-flight (Q-TOF) equipped with a low-voltage digital waveform driver to navigate the filtering quadrupole in the HZs. We demonstrate that a digital quadrupole operating in zones (3,1) and (3,2) achieves a narrowed isolation window for analytes up to 690 and 1543 <i>m</i>/<i>z</i>, respectively, compared to conventional operational modes. Experimental trends suggest this performance to continue to approximately 900 <i>m</i>/<i>z</i> in zone (3,1) and 2500 <i>m</i>/<i>z</i> in zone (3,2). These narrowed isolation windows were utilized to generate MS/MS data from a mixture of <i>m</i>-hydroxybenzoylecgonine (306 <i>m</i>/<i>z</i>) and cocaine-d3 (307 <i>m</i>/<i>z</i>), which exhibited minimal chimeric nature without the use of deconvolution software. The development of hybrid systems capable of both sine and digital operation provides a mechanism to maximize selectivity of the HZs while maintaining the benefits of traditional modes of operation. Minimizing chimeric interferences is of particular importance, and the narrow isolation widths afforded by HZs are readily accessible using tandem instruments for analytes <2500 <i>m</i>/<i>z</i>.

Authors

Institutions

Publication Details

Journal
Journal of the American Society for Mass Spectrometry
Published
2025-12-20
DOI
https://doi.org/10.1021/jasms.5c00312
Primary Topic
Molecular Spectroscopy and Structure
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Accessing Higher Order Mathieu Space Stability Zones to Narrow Isolation Widths Using Digital Quadrupole Time-of-Flight Mass Spectrometry

Shane E. Tichy, Elizabeth Groetsema, Brian H. Clowers, Peter T. A. Reilly et al.
Journal of the American Society for Mass Spectrometry
Molecular Spectroscopy and Structure
article

Accessing Higher Order Mathieu Space Stability Zones to Narrow Isolation Widths Using Digital Quadrupole Time-of-Flight Mass Spectrometry

Shane E. Tichy, Elizabeth Groetsema, Brian H. Clowers, Peter T. A. Reilly, Adam P. Huntley
article en

Abstract

Digital quadrupoles are driven with rectangular RF waveforms and, through duty cycle control, can access higher-order Mathieu space stability zones (HZs) using comparatively low voltages. Analytically, accessing these zones remains attractive as HZs exhibit higher resolving powers ((<i>m</i>/<i>z</i>)/(Δ<i>m</i>/<i>z</i>)) compared to the conventional quadrupole operation in zone (1,1). Presented here is a modification of a commercial quadrupole time-of-flight (Q-TOF) equipped with a low-voltage digital waveform driver to navigate the filtering quadrupole in the HZs. We demonstrate that a digital quadrupole operating in zones (3,1) and (3,2) achieves a narrowed isolation window for analytes up to 690 and 1543 <i>m</i>/<i>z</i>, respectively, compared to conventional operational modes. Experimental trends suggest this performance to continue to approximately 900 <i>m</i>/<i>z</i> in zone (3,1) and 2500 <i>m</i>/<i>z</i> in zone (3,2). These narrowed isolation windows were utilized to generate MS/MS data from a mixture of <i>m</i>-hydroxybenzoylecgonine (306 <i>m</i>/<i>z</i>) and cocaine-d3 (307 <i>m</i>/<i>z</i>), which exhibited minimal chimeric nature without the use of deconvolution software. The development of hybrid systems capable of both sine and digital operation provides a mechanism to maximize selectivity of the HZs while maintaining the benefits of traditional modes of operation. Minimizing chimeric interferences is of particular importance, and the narrow isolation widths afforded by HZs are readily accessible using tandem instruments for analytes <2500 <i>m</i>/<i>z</i>.

Journal of the American Society for Mass SpectrometryVol. 37(1)
Agilent Technologies (United States) (US), Agilent Technologies (Germany) (DE), Washington State University (US)
Openalex Percentile: Top 63%
Molecular Spectroscopy and Structure
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.