Noninvasive blood pressure pumps are recommended as the primary form of acquiring patient blood pressure to test for blood poisoning. The Philips IntelliVue Patient Monitor MP5, the Philips SureSigns VS4/VS3 Vital Signs Monitor, the Philips EarlyVue VS30 Vital Signs Monitor, the Philips IntelliVue Patent Monitor MP2/X2, the Philips IntelliVue Patent Monitor MMS, and the Philips IntelliVue Patent Monitor MP3/X3 were tested for accuracy. The control condition for hypotension was 95/55 (68) mmHg, normal blood pressure was 120/80 (93) mmHg, and hypertension was 155/85 (108) mmHg. The order of values is arrayed according to diastolic, systolic, and mean arterial pressure. The results displayed no observable difference between error or uncertainty. The average relative uncertainty values were identical for calibrated and uncalibrated devices. These values included 0.32%, 0.38%, and 0.51% for hypotension, normal blood pressure, and hypertension respectively. This result is significantly smaller than the blood pressure values. Also, devices with the greatest deviation in variance were the Philips IntelliVue Patient Monitor MP5 and X2, which exhibited the maximum standard deviation in seven out of the nine cases. The device with the smallest deviation in variance was the Philips IntelliVue Patient Monitor VS3/4 with a minimum variance of zero. Further experimentation is encouraged to collect a larger sample size to complete hypothesis testing and compare it to the industrial quality control of Philips.

Authors

Publication Details

Journal
Journal of Purdue Undergraduate Research
Published
2025-12-19
DOI
https://doi.org/10.7771/2158-4052.1738
Primary Topic
Blood Pressure and Hypertension Studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Implications of Noninvasive Blood Pressure Pumps on Mean Arterial Pressure

Stephen E. Miller
Journal of Purdue Undergraduate Research
Blood Pressure and Hypertension Studies
article

Implications of Noninvasive Blood Pressure Pumps on Mean Arterial Pressure

Stephen E. Miller
article en

Abstract

Noninvasive blood pressure pumps are recommended as the primary form of acquiring patient blood pressure to test for blood poisoning. The Philips IntelliVue Patient Monitor MP5, the Philips SureSigns VS4/VS3 Vital Signs Monitor, the Philips EarlyVue VS30 Vital Signs Monitor, the Philips IntelliVue Patent Monitor MP2/X2, the Philips IntelliVue Patent Monitor MMS, and the Philips IntelliVue Patent Monitor MP3/X3 were tested for accuracy. The control condition for hypotension was 95/55 (68) mmHg, normal blood pressure was 120/80 (93) mmHg, and hypertension was 155/85 (108) mmHg. The order of values is arrayed according to diastolic, systolic, and mean arterial pressure. The results displayed no observable difference between error or uncertainty. The average relative uncertainty values were identical for calibrated and uncalibrated devices. These values included 0.32%, 0.38%, and 0.51% for hypotension, normal blood pressure, and hypertension respectively. This result is significantly smaller than the blood pressure values. Also, devices with the greatest deviation in variance were the Philips IntelliVue Patient Monitor MP5 and X2, which exhibited the maximum standard deviation in seven out of the nine cases. The device with the smallest deviation in variance was the Philips IntelliVue Patient Monitor VS3/4 with a minimum variance of zero. Further experimentation is encouraged to collect a larger sample size to complete hypothesis testing and compare it to the industrial quality control of Philips.

Journal of Purdue Undergraduate ResearchVol. 15(1)
Openalex Percentile: Top 42%
Blood Pressure and Hypertension Studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.