Evaluation of lead equivalence and integrity of personal radiation protective equipment using multiple imaging modalities

Personal radiation protective equipment (PRPE) play a critical role in minimising occupational exposure to ionising radiation in medical imaging, nuclear medicine, interventional procedures and related fields. Over time, repeated use, mechanical stress, and improper handling can degrade their protective performance, making regular inspection essential to ensure continued radiation safety. This study aimed to compare the performance of multiple imaging modalities in evaluating the integrity and lead equivalence of PRPE, to identify the most accurate and practical methods for quality assurance programmes. Over a four-year period, 1063 PRPE items from 20 manufacturers were evaluated. Each item underwent visual and tactile inspection, followed by imaging using radiography, fluoroscopy, computed tomography (CT), and a dedicated PRPE testing system (FLOWD 8020). Lead equivalence was determined by comparison with certified reference lead foils using both x-ray quality assurance dosimeter and image-based methods. Integrity was categorised on a four-level scale according to defect size and location, and attenuation performance was assessed relative to manufacturers' nominal specifications. Of all PRPE items tested, 88.8% were defect-free, while 11.2% showed cracks or tears cracks or tears of varying extent. Lead equivalence results revealed that 74.3% met specifications within ±5%, 13.8% deviated by 5%-10%, and 11.9% by more than 10%. All imaging modalities demonstrated comparable accuracy, though the dedicated screening system offered clear advantages in workflow efficiency, full-apron coverage, automated reporting, and low stray dose (<0.3<i>μ</i>Sv h<sup>-1</sup>at 1 m). Routine PRPE inspection is essential for maintaining occupational radiation safety. Standardised testing protocols and inspection intervals are recommended to ensure consistent and traceable quality assurance practices across institutions.

Authors

Institutions

Publication Details

Journal
Journal of Radiological Protection
Published
2025-12-19
DOI
https://doi.org/10.1088/1361-6498/ae2f6b
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of lead equivalence and integrity of personal radiation protective equipment using multiple imaging modalities

Kirill Skovorodko, Antonio Jreije, Leonid Krynke, Milda Bareikė et al.
Journal of Radiological Protection
Radiation Dose and Imaging
article

Evaluation of lead equivalence and integrity of personal radiation protective equipment using multiple imaging modalities

Kirill Skovorodko, Antonio Jreije, Leonid Krynke, Milda Bareikė, Kristina Kristinaitytė, Inga Andriulevičiūtė, Birutė Gricienė
article en

Abstract

Personal radiation protective equipment (PRPE) play a critical role in minimising occupational exposure to ionising radiation in medical imaging, nuclear medicine, interventional procedures and related fields. Over time, repeated use, mechanical stress, and improper handling can degrade their protective performance, making regular inspection essential to ensure continued radiation safety. This study aimed to compare the performance of multiple imaging modalities in evaluating the integrity and lead equivalence of PRPE, to identify the most accurate and practical methods for quality assurance programmes. Over a four-year period, 1063 PRPE items from 20 manufacturers were evaluated. Each item underwent visual and tactile inspection, followed by imaging using radiography, fluoroscopy, computed tomography (CT), and a dedicated PRPE testing system (FLOWD 8020). Lead equivalence was determined by comparison with certified reference lead foils using both x-ray quality assurance dosimeter and image-based methods. Integrity was categorised on a four-level scale according to defect size and location, and attenuation performance was assessed relative to manufacturers' nominal specifications. Of all PRPE items tested, 88.8% were defect-free, while 11.2% showed cracks or tears cracks or tears of varying extent. Lead equivalence results revealed that 74.3% met specifications within ±5%, 13.8% deviated by 5%-10%, and 11.9% by more than 10%. All imaging modalities demonstrated comparable accuracy, though the dedicated screening system offered clear advantages in workflow efficiency, full-apron coverage, automated reporting, and low stray dose (<0.3<i>μ</i>Sv h<sup>-1</sup>at 1 m). Routine PRPE inspection is essential for maintaining occupational radiation safety. Standardised testing protocols and inspection intervals are recommended to ensure consistent and traceable quality assurance practices across institutions.

Journal of Radiological ProtectionVol. 46(1)
Vilnius University (LT), Vilnius University of Applied Sciences (LT), Vilnius University Hospital Santariskiu Klinikos (LT)
Openalex Percentile: Top 44%
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.