Effect on the full peak efficiency of HPGe detectors against different sample densities

The study aims to investigate the density effect on the full energy peak efficiency (FEPE) of five identical certified standard volume sources of the same geometry, radioisotopes, and activity but having different densities (0.7-1.6 g.cm-3) experimentally using a P-type HPGe detector of 100% relative efficiency and theoretically using ANGLE software. The count rate of γ-energy lines of the γ-sources 137Cs and 60Co were measured experimentally. Four methods were used to discuss the density effect of these sources on the full energy peak efficiency by showing the FEPE deviation percentages between the calculated and certified sample-specific activities. The methods showed that while ignoring the density of the source, the FEPE deviation percentages were very high , with an average of 14.025%. Considering the sources' density using either ANGLE-3 software or relative methods, the FEPE deviation percentages were very low, with acceptable averages of 2.13%.

Authors

Institutions

Publication Details

Journal
International Journal of Power and Energy Conversion
Published
2025-12-19
DOI
https://doi.org/10.1504/ijpec.2026.150621
Primary Topic
Radioactivity and Radon Measurements
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Effect on the full peak efficiency of HPGe detectors against different sample densities

Wael El Gammal, Islam M. Nabil, Y. Y. Ebaid, K.M. El Kourghly et al.
International Journal of Power and Energy Conversion
Radioactivity and Radon Measurements
article

Effect on the full peak efficiency of HPGe detectors against different sample densities

Wael El Gammal, Islam M. Nabil, Y. Y. Ebaid, K.M. El Kourghly, Y. Mohamed, Omar Abu Bakr
article en

Abstract

The study aims to investigate the density effect on the full energy peak efficiency (FEPE) of five identical certified standard volume sources of the same geometry, radioisotopes, and activity but having different densities (0.7-1.6 g.cm-3) experimentally using a P-type HPGe detector of 100% relative efficiency and theoretically using ANGLE software. The count rate of γ-energy lines of the γ-sources 137Cs and 60Co were measured experimentally. Four methods were used to discuss the density effect of these sources on the full energy peak efficiency by showing the FEPE deviation percentages between the calculated and certified sample-specific activities. The methods showed that while ignoring the density of the source, the FEPE deviation percentages were very high , with an average of 14.025%. Considering the sources' density using either ANGLE-3 software or relative methods, the FEPE deviation percentages were very low, with acceptable averages of 2.13%.

International Journal of Power and Energy ConversionVol. 17(1)
Military Technical College (EG), Fayoum University (EG)
Openalex Percentile: Top 46%
Radioactivity and Radon Measurements
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.