A CADAVER BIOMECHANICAL COMPARISON OF THE PLATE CONSTRUCT FOR OPEN LAPIDUS FUSION VERSUS PERCUTANEOUS LAPIDUS FUSION USING SUBCHONDRAL FIXATION

Introduction: Tarsometatarsal (TMT) fusions traditionally use bridge plates or compression screws through open incisions to correct deformity and achieve bone healing. A new percutaneous technique can remove cartilage, preserve the subchondral bone and transfix the fusion site. Intraosseus, headless, chamfered, full-thread, non-variable pitch screws are used to transfix all available bone. This study compared the biomechanical properties of the plate construct versus the percutaneous construct. Method: TMT joints of six cadavers (12 limbs) were stabilized with a locking plate and 3.5 mm compression screw on one limb, and with a percutaneous Lapidus construct on the matched limb. An extensometer was placed on the plantar side of the TMT. The first metatarsal was point loaded cyclically from 9 to 90 N at 60 mm from the TMT joint at 3 Hz. Testing was stopped if the extensometer reached 7 mm of plantar gap, or when 250,000 cycles were reached. Specimens were then statically loaded to failure. Results: Percutaneous screw specimens reached more cycles to failure at 226,000 ± 58,000 versus 30,000 ± 52,000 for the open plate construct (p<.001). Plantar gap was higher in the open plate construct at all cycle counts from 10 to 10,000 cycles (>4 mm) versus the percutaneous group (<1 mm). Maximum load was higher in the percutaneous group (343.3 ± 92.8 N versus 247.3 ± 28.3 N) (p<.05). Stiffness of the percutaneous group was higher (40.9 ± 13.4 N/mm versus 16.0 ± 5.3 N/mm) (p<.01). No differences were found in bone quality testing. A Cox proportional hazard model identified the mode of fixation as the only significant covariant (p<.05) in predicting cycles to failure. Conclusion: The percutaneous subchondral fixation fusion construct with full-thread, non-variable pitch, intraosseous screws create a stiffer and stronger fixation construct than dorsal plates and cross-screws.

Authors

Institutions

Publication Details

Journal
Orthopaedic Proceedings
Published
2025-12-19
DOI
https://doi.org/10.1302/1358-992x.2025.17.017
Primary Topic
Foot and Ankle Surgery
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A CADAVER BIOMECHANICAL COMPARISON OF THE PLATE CONSTRUCT FOR OPEN LAPIDUS FUSION VERSUS PERCUTANEOUS LAPIDUS FUSION USING SUBCHONDRAL FIXATION

Björn Thorén, Douglas Linton, Nathan R. Webb, Alastair S. E. Younger et al.
Orthopaedic Proceedings
Foot and Ankle Surgery
article

A CADAVER BIOMECHANICAL COMPARISON OF THE PLATE CONSTRUCT FOR OPEN LAPIDUS FUSION VERSUS PERCUTANEOUS LAPIDUS FUSION USING SUBCHONDRAL FIXATION

Björn Thorén, Douglas Linton, Nathan R. Webb, Alastair S. E. Younger, Stephen Steinlauf
article en

Abstract

Introduction: Tarsometatarsal (TMT) fusions traditionally use bridge plates or compression screws through open incisions to correct deformity and achieve bone healing. A new percutaneous technique can remove cartilage, preserve the subchondral bone and transfix the fusion site. Intraosseus, headless, chamfered, full-thread, non-variable pitch screws are used to transfix all available bone. This study compared the biomechanical properties of the plate construct versus the percutaneous construct. Method: TMT joints of six cadavers (12 limbs) were stabilized with a locking plate and 3.5 mm compression screw on one limb, and with a percutaneous Lapidus construct on the matched limb. An extensometer was placed on the plantar side of the TMT. The first metatarsal was point loaded cyclically from 9 to 90 N at 60 mm from the TMT joint at 3 Hz. Testing was stopped if the extensometer reached 7 mm of plantar gap, or when 250,000 cycles were reached. Specimens were then statically loaded to failure. Results: Percutaneous screw specimens reached more cycles to failure at 226,000 ± 58,000 versus 30,000 ± 52,000 for the open plate construct (p<.001). Plantar gap was higher in the open plate construct at all cycle counts from 10 to 10,000 cycles (>4 mm) versus the percutaneous group (<1 mm). Maximum load was higher in the percutaneous group (343.3 ± 92.8 N versus 247.3 ± 28.3 N) (p<.05). Stiffness of the percutaneous group was higher (40.9 ± 13.4 N/mm versus 16.0 ± 5.3 N/mm) (p<.01). No differences were found in bone quality testing. A Cox proportional hazard model identified the mode of fixation as the only significant covariant (p<.05) in predicting cycles to failure. Conclusion: The percutaneous subchondral fixation fusion construct with full-thread, non-variable pitch, intraosseous screws create a stiffer and stronger fixation construct than dorsal plates and cross-screws.

Orthopaedic ProceedingsVol. 107-B(SUPP_17)
University of British Columbia (CA), Stryker (United States) (US)
Openalex Percentile: Top 49%
Foot and Ankle Surgery
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.