Patient-specific guides for minimally invasive sacroiliac joint fusion revision surgery

• Novel patient-specific guide for minimally invasive SI joint fusion revision surgery. • Guide positioning is based on previously placed implants; no bone seating is required. • Non-parallel implant positioning beyond traditional parallel configuration. • Non-parallel is favorable, especially if there is limited space in the SI joint. • Accuracy was evaluated using 3D positional and angular measurements. Implant loosening or persistent pain after sacroiliac joint fusion or stabilization surgery may require revision surgery. Accurate and safe implant placement in minimally invasive sacroiliac joint fusion (SIJF) revision surgery is challenging due to limited visibility and restricted space for new implants. A patient-specific guide (PSG) allows customization of implant positioning beyond the traditional parallel configuration generally used in SIJF but has only been applied in open procedures requiring bone exposure. This study presents the implementation and evaluation of a novel PSG for minimally invasive SIJF revision surgery in patients with prior SI fusion or stabilization surgery. A retrospective cohort study was conducted to evaluate implant placement accuracy using a novel PSG. The guide design was iteratively refined to enhance rigidity and placement accuracy. Postoperative implant positions were compared to the planned positions using 3D deviation analysis. Furthermore, implant malpositioning and other complications were monitored. Fourteen SIJF revisions were performed, comprising a total of 23 newly placed implants. Implant positioning using PSGs was feasible in all cases, and all implants were placed without malpositioning complications. Accuracy analysis showed a median 3D total positional deviation of 5.3 mm (Q1: 4.8 mm, Q3: 8.3 mm) and a median total 3D angle of 3.7° (Q1: 2.4°, Q3: 5.2°). The study showed that the developed PSG design for use in minimally invasive SIJF revision surgery is feasible with what is considered a clinically acceptable level of accuracy.

Authors

Institutions

Publication Details

Journal
Annals of 3D Printed Medicine
Published
2025-12-19
DOI
https://doi.org/10.1016/j.stlm.2025.100225
Primary Topic
Spine and Intervertebral Disc Pathology
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Patient-specific guides for minimally invasive sacroiliac joint fusion revision surgery

Maaike A. Koenrades, Jorm M. Nellensteijn, Gabriëlle J. M. Tuijthof, Edsko E.G. Hekman et al.
Annals of 3D Printed Medicine
Spine and Intervertebral Disc Pathology
article

Patient-specific guides for minimally invasive sacroiliac joint fusion revision surgery

Maaike A. Koenrades, Jorm M. Nellensteijn, Gabriëlle J. M. Tuijthof, Edsko E.G. Hekman, Nick Kampkuiper, Femke F. Schröder, Danique Smelt, Maranda Haenen
article en

Abstract

• Novel patient-specific guide for minimally invasive SI joint fusion revision surgery. • Guide positioning is based on previously placed implants; no bone seating is required. • Non-parallel implant positioning beyond traditional parallel configuration. • Non-parallel is favorable, especially if there is limited space in the SI joint. • Accuracy was evaluated using 3D positional and angular measurements. Implant loosening or persistent pain after sacroiliac joint fusion or stabilization surgery may require revision surgery. Accurate and safe implant placement in minimally invasive sacroiliac joint fusion (SIJF) revision surgery is challenging due to limited visibility and restricted space for new implants. A patient-specific guide (PSG) allows customization of implant positioning beyond the traditional parallel configuration generally used in SIJF but has only been applied in open procedures requiring bone exposure. This study presents the implementation and evaluation of a novel PSG for minimally invasive SIJF revision surgery in patients with prior SI fusion or stabilization surgery. A retrospective cohort study was conducted to evaluate implant placement accuracy using a novel PSG. The guide design was iteratively refined to enhance rigidity and placement accuracy. Postoperative implant positions were compared to the planned positions using 3D deviation analysis. Furthermore, implant malpositioning and other complications were monitored. Fourteen SIJF revisions were performed, comprising a total of 23 newly placed implants. Implant positioning using PSGs was feasible in all cases, and all implants were placed without malpositioning complications. Accuracy analysis showed a median 3D total positional deviation of 5.3 mm (Q1: 4.8 mm, Q3: 8.3 mm) and a median total 3D angle of 3.7° (Q1: 2.4°, Q3: 5.2°). The study showed that the developed PSG design for use in minimally invasive SIJF revision surgery is feasible with what is considered a clinically acceptable level of accuracy.

Annals of 3D Printed MedicineVol. 21
Medisch Spectrum Twente (NL), University of Twente (NL)
Openalex Percentile: Top 43%
Spine and Intervertebral Disc Pathology
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.