Beam commissioning of a new compact scanned proton therapy system with four different dose calculation algorithms
All calculated doses of the box plans were in good agreement with the measurements, except for the RayStation PBA. For clinical use, RayStation MC is the preferred option for both homogeneous and heterogeneous cases.