Monte Carlo-based PET range verification for proton therapy in breast cancer patients with silicone implants: material assignment impact analysis

<i>Objective.</i>To validate a Monte Carlo (MC)-based offline positron emission tomography (PET) range verification framework and quantitatively evaluate range uncertainties associated with material assignment strategies, specifically focusing on the clinical practice of overriding silicone implants as adipose tissue in breast proton therapy.<i>Approach.</i>A FLUKA-based MC model was developed. Validation followed a stepwise program using (1) a PMMA phantom, (2) a homogeneous silicone block, and (3) a composite prosthesis-PMMA setup to verify intrinsic precision and material-specific accuracy. Additionally, a cohort of eight breast cancer patients (three with implants) was included for clinical demonstration. Range deviations were quantified by measuring the depth difference at the 50% distal activity falloff (ΔR50) between simulated and measured PET images. For silicone scenarios, material assignment impact was analyzed by overriding the implant as silicone, adipose tissue, or applying no override.<i>Main results.</i>The MC model demonstrated intrinsic millimeter-level accuracy. In homogeneous silicone phantom experiments, defining the material as silicone yielded the highest agreement with measurements (ΔR50: -0.05-0.88 mm). Reassigning silicone as adipose tissue introduced a systematic deviation but remained within acceptable limits (-1.24--0.83 mm), whereas applying no override resulted in the largest errors (-3.25--2.48 mm). Composite phantom results followed a similar trend. In clinical cases, the method successfully verified range delivery, although positioning uncertainties contributed to larger inter-field variability.<i>Significance.</i>This study validates an MC-based PET verification framework with 2 mm intrinsic accuracy, establishing it as a reliable tool for silicone-implanted breast cancer patients. The results quantitatively confirm that overriding silicone as adipose tissue keeps range uncertainty within 2 mm despite distinct elemental compositions, whereas failing to override the material leads to errors exceeding 3 mm.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2025-12-18
DOI
https://doi.org/10.1088/1361-6560/ae2f17
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Monte Carlo-based PET range verification for proton therapy in breast cancer patients with silicone implants: material assignment impact analysis

Jiangang Zhang, Yinxiangzi Sheng, Jingyi Cheng, Qing Zhang et al.
Physics in Medicine and Biology
Medical Imaging Techniques and Applications
article

Monte Carlo-based PET range verification for proton therapy in breast cancer patients with silicone implants: material assignment impact analysis

Jiangang Zhang, Yinxiangzi Sheng, Jingyi Cheng, Qing Zhang, Jiayi Guo, Yanzhao Wang
article en

Abstract

<i>Objective.</i>To validate a Monte Carlo (MC)-based offline positron emission tomography (PET) range verification framework and quantitatively evaluate range uncertainties associated with material assignment strategies, specifically focusing on the clinical practice of overriding silicone implants as adipose tissue in breast proton therapy.<i>Approach.</i>A FLUKA-based MC model was developed. Validation followed a stepwise program using (1) a PMMA phantom, (2) a homogeneous silicone block, and (3) a composite prosthesis-PMMA setup to verify intrinsic precision and material-specific accuracy. Additionally, a cohort of eight breast cancer patients (three with implants) was included for clinical demonstration. Range deviations were quantified by measuring the depth difference at the 50% distal activity falloff (ΔR50) between simulated and measured PET images. For silicone scenarios, material assignment impact was analyzed by overriding the implant as silicone, adipose tissue, or applying no override.<i>Main results.</i>The MC model demonstrated intrinsic millimeter-level accuracy. In homogeneous silicone phantom experiments, defining the material as silicone yielded the highest agreement with measurements (ΔR50: -0.05-0.88 mm). Reassigning silicone as adipose tissue introduced a systematic deviation but remained within acceptable limits (-1.24--0.83 mm), whereas applying no override resulted in the largest errors (-3.25--2.48 mm). Composite phantom results followed a similar trend. In clinical cases, the method successfully verified range delivery, although positioning uncertainties contributed to larger inter-field variability.<i>Significance.</i>This study validates an MC-based PET verification framework with 2 mm intrinsic accuracy, establishing it as a reliable tool for silicone-implanted breast cancer patients. The results quantitatively confirm that overriding silicone as adipose tissue keeps range uncertainty within 2 mm despite distinct elemental compositions, whereas failing to override the material leads to errors exceeding 3 mm.

Physics in Medicine and BiologyVol. 71(1)
Sun Yat-sen University (CN), Donghua University (CN), Shanghai Medical Information Center (CN), Fudan University Shanghai Cancer Center (CN), Shanghai Genon Biological Products (China) (CN), Center for High Pressure Science & Technology Advanced Research (CN), Sun Yat-sen University Cancer Center (CN)
Openalex Percentile: Top 44%
Medical Imaging Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.