Mechanically-foldable axial flow blood pump: response-surface-based structural optimization and hemolytic performance evaluation

We designed and optimized a foldable axial-flow blood pump for transaortic implantation. The optimized configuration demonstrates a 6.9% head improvement to 2.346 m and 17.9% HI reduction to <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML"><mml:mrow><mml:mn>1.081</mml:mn> <mml:mo>×</mml:mo> <mml:mn>1</mml:mn> <mml:msup><mml:mrow><mml:mn>0</mml:mn></mml:mrow> <mml:mrow><mml:mo>-</mml:mo> <mml:mn>2</mml:mn></mml:mrow> </mml:msup> <mml:mi>%</mml:mi></mml:mrow> </mml:math> . Furthermore, increasing tip clearance from 0.2 mm to 0.4 mm provides additional hemolysis reduction without excessively compromising the pressure head.

Authors

Institutions

Publication Details

Journal
Frontiers in Physiology
Published
2025-12-18
DOI
https://doi.org/10.3389/fphys.2025.1632333
Citations
1
Primary Topic
Mechanical Circulatory Support Devices
Type
article
Field-Weighted Citation Impact
0.67
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Mechanically-foldable axial flow blood pump: response-surface-based structural optimization and hemolytic performance evaluation

Jing Teng, Jianan Cheng, T. Ji
1 citations
Frontiers in Physiology
Mechanical Circulatory Support Devices
0.67
article

Mechanically-foldable axial flow blood pump: response-surface-based structural optimization and hemolytic performance evaluation

Jing Teng, Jianan Cheng, T. Ji
article en
1 citations

Abstract

We designed and optimized a foldable axial-flow blood pump for transaortic implantation. The optimized configuration demonstrates a 6.9% head improvement to 2.346 m and 17.9% HI reduction to <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML"><mml:mrow><mml:mn>1.081</mml:mn> <mml:mo>×</mml:mo> <mml:mn>1</mml:mn> <mml:msup><mml:mrow><mml:mn>0</mml:mn></mml:mrow> <mml:mrow><mml:mo>-</mml:mo> <mml:mn>2</mml:mn></mml:mrow> </mml:msup> <mml:mi>%</mml:mi></mml:mrow> </mml:math> . Furthermore, increasing tip clearance from 0.2 mm to 0.4 mm provides additional hemolysis reduction without excessively compromising the pressure head.

Frontiers in PhysiologyVol. 16
Jiangsu University (CN)
Openalex Percentile: Top 24%
Mechanical Circulatory Support Devices
0.67
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.