Enhanced fabrication of dual-compartment artificial cornea C-Clear <i>via</i> precision moulding and continuous polymerisation: biocompatibility and functional efficacy in rabbit model

This study introduces C-Clear, a novel artificial cornea based on a HEMA/MMA-based copolymer, developed through continuous polymerisation and moulding. C-Clear comprises a transparent optical core and porous support skirt, specifically designed to enhance tissue integration and minimise inflammatory responses. <i>In vitro</i> evaluations demonstrated excellent biocompatibility, characterised by high levels of cell adhesion and proliferation, while <i>in vivo</i> assessments using a rat subcutaneous model confirmed successful integration and biocompatibility. Furthermore, a 24-week corneal implantation study in rabbits validated the stability, safety, and functional potential of C-Clear. Serial ophthalmic examinations during this study period showed no significant progression of neovascularisation or inflammation. Histological analyses revealed exceptional optical clarity, robust integration with surrounding tissues, and an absence of notable foreign body responses. The implant achieved a retention rate of 75% over the 24 weeks, further highlighting its reliability. The custom-designed mould and continuous polymerisation process enabled the fabrication of C-Clear with superior structural stability, biocompatibility, and therapeutic efficacy. These findings highlight C-Clear as a significant advancement in artificial corneal development, addressing the global shortage of donor corneas and offering a promising solution for treating corneal blindness.

Authors

Institutions

Publication Details

Journal
Journal of Materials Chemistry B
Published
2025-12-18
DOI
https://doi.org/10.1039/d5tb01842k
Primary Topic
Corneal Surgery and Treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Enhanced fabrication of dual-compartment artificial cornea C-Clear via precision moulding and continuous polymerisation: biocompatibility and functional efficacy in rabbit model

Eun‐Ji Choi, Joong-Hyun Kim, Tae Hyun Kim, Jae Hong Park et al.
Journal of Materials Chemistry B
Corneal Surgery and Treatments
article

Enhanced fabrication of dual-compartment artificial cornea C-Clear via precision moulding and continuous polymerisation: biocompatibility and functional efficacy in rabbit model

Eun‐Ji Choi, Joong-Hyun Kim, Tae Hyun Kim, Jae Hong Park, Chul Min Kim, Do-Sun Jeong
article en

Abstract

This study introduces C-Clear, a novel artificial cornea based on a HEMA/MMA-based copolymer, developed through continuous polymerisation and moulding. C-Clear comprises a transparent optical core and porous support skirt, specifically designed to enhance tissue integration and minimise inflammatory responses. <i>In vitro</i> evaluations demonstrated excellent biocompatibility, characterised by high levels of cell adhesion and proliferation, while <i>in vivo</i> assessments using a rat subcutaneous model confirmed successful integration and biocompatibility. Furthermore, a 24-week corneal implantation study in rabbits validated the stability, safety, and functional potential of C-Clear. Serial ophthalmic examinations during this study period showed no significant progression of neovascularisation or inflammation. Histological analyses revealed exceptional optical clarity, robust integration with surrounding tissues, and an absence of notable foreign body responses. The implant achieved a retention rate of 75% over the 24 weeks, further highlighting its reliability. The custom-designed mould and continuous polymerisation process enabled the fabrication of C-Clear with superior structural stability, biocompatibility, and therapeutic efficacy. These findings highlight C-Clear as a significant advancement in artificial corneal development, addressing the global shortage of donor corneas and offering a promising solution for treating corneal blindness.

Journal of Materials Chemistry BVol. 14(3)
Gyeongsang National University (KR), Osong Medical Innovation Foundation (KR)
Openalex Percentile: Top 44%
Corneal Surgery and Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.