Spin-Exchange Relaxation-Free Optically Pumped Magnetometers (SERF-OPMs) are increasingly used in multichannel biomagnetic sensing, yet their timing performance remains poorly characterized. This contribution presents the first cross-platform study of time delay, group delay, intra-channel variability, and settling time across four commercial SERF-OPM systems: Neuro-1 and QZFM Gen.2 (QuSpin Inc.), and HEDscan and FLv2 (FieldLine Inc.). Measurements were performed inside a magnetically shielded room (BMSR-2.1, PTB, Berlin) using an already introduced test bench. Results show frequency-dependent delays of 1-10 ms, intra-channel spreads up to +-1 ms, group delays between 1-15 ms, and settling times of 2-55 ms. Clear differences in manufacturer strategies were observed: QuSpin minimizes intra-channel variability through digital delay equalization, whereas FieldLine employs per-sensor calibration to optimize bandwidth and phase matching. In all systems, the time delay deviation between channels is in the sub-millisecond range in the 20-140 Hz band, which is sufficient for magnetoencephalography source localization. However, longer settling times in some platforms limit performance for rapid stimulation protocols. These findings provide the first vendor-comprehensive benchmarks for timing parameters in commercial SERF-OPM systems. They highlight the trade-offs between bandwidth, delay, and calibration strategies, and underscore the need for rigorous timing characterization to ensure waveform fidelity. The results are directly relevant to applications such as stimulation-evoked responses, brain-computer interfaces, and closed-loop neuromodulation.

Authors

Publication Details

Journal
Measurement
Published
2025-12-17
DOI
https://doi.org/10.1016/j.measurement.2025.120140
Primary Topic
Atomic and Subatomic Physics Research
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Characterizing timing parameters in commercial SERF-OPM multichannel systems for biomagnetic field sensing

Stephen E. Robinson, Michael Höft, Surjo R. Soekadar, Eric Elzenheimer et al.
Measurement
Atomic and Subatomic Physics Research
article

Characterizing timing parameters in commercial SERF-OPM multichannel systems for biomagnetic field sensing

Stephen E. Robinson, Michael Höft, Surjo R. Soekadar, Eric Elzenheimer, S. Knappe-Grüneberg, Paul Anders, Robert Rieger, Jan Zerfowski, Hartmut Matz
article en

Abstract

Spin-Exchange Relaxation-Free Optically Pumped Magnetometers (SERF-OPMs) are increasingly used in multichannel biomagnetic sensing, yet their timing performance remains poorly characterized. This contribution presents the first cross-platform study of time delay, group delay, intra-channel variability, and settling time across four commercial SERF-OPM systems: Neuro-1 and QZFM Gen.2 (QuSpin Inc.), and HEDscan and FLv2 (FieldLine Inc.). Measurements were performed inside a magnetically shielded room (BMSR-2.1, PTB, Berlin) using an already introduced test bench. Results show frequency-dependent delays of 1-10 ms, intra-channel spreads up to +-1 ms, group delays between 1-15 ms, and settling times of 2-55 ms. Clear differences in manufacturer strategies were observed: QuSpin minimizes intra-channel variability through digital delay equalization, whereas FieldLine employs per-sensor calibration to optimize bandwidth and phase matching. In all systems, the time delay deviation between channels is in the sub-millisecond range in the 20-140 Hz band, which is sufficient for magnetoencephalography source localization. However, longer settling times in some platforms limit performance for rapid stimulation protocols. These findings provide the first vendor-comprehensive benchmarks for timing parameters in commercial SERF-OPM systems. They highlight the trade-offs between bandwidth, delay, and calibration strategies, and underscore the need for rigorous timing characterization to ensure waveform fidelity. The results are directly relevant to applications such as stimulation-evoked responses, brain-computer interfaces, and closed-loop neuromodulation.

MeasurementVol. 263
Openalex Percentile: Top 67%
Atomic and Subatomic Physics Research
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.