STEPC: A Pixel-wise Nonuniformity Correction Framework for Photon-Counting CT in Multi-material Imaging Scenarios

Photon-counting computed tomography (PCCT) has demonstrated significant advancements in recent years; however, pixel-wise detector response nonuniformity remains a key challenge, frequently manifesting as ring artifacts in reconstructed images. Existing correction methods exhibit limited generalizability in complex multi-material scenarios, such as contrast-enhanced imaging. This study introduces a Signal-to-Uniformity Error Polynomial Calibration (STEPC) framework to address this issue. STEPC first fits multi-energy projections using a 2D polynomial surface to generate ideal references, then applies a nonlinear multi-energy polynomial model to predict and correct pixel-wise nonuniformity errors. The model is calibrated using homogeneous slab phantoms of different materials, including PMMA, aluminum, and iodinated contrast agents, enabling correction for both non-contrast and contrast-enhanced imaging. Experiments were performed on a custom Micro-PCCT system with phantoms and mouse. Correction performance of STEPC was evaluated using the mean local standard deviation (MLSD) in the projection domain and the ring artifact deviation (RAD) on the reconstructed images. STEPC consistently outperformed existing correction methods in both non-contrast and contrast-enhanced scenarios. It achieved the lowest MLSD and RAD for both phantoms and mouse scans. These results indicate that STEPC provides a robust and practical solution for correcting detector nonuniformity in multi-material PCCT imaging, witch position it as a promising general-purpose calibration framework for photon-counting CT systems.

Authors

Publication Details

Journal
ArXiv.org
Published
2025-07-20
Citations
1
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.67
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

STEPC: A Pixel-wise Nonuniformity Correction Framework for Photon-Counting CT in Multi-material Imaging Scenarios

Enze Zhou, Shaoyang Wang, Qian Liu, Wenting Xu et al.
1 citations
ArXiv.org
Advanced X-ray and CT Imaging
0.67
article

STEPC: A Pixel-wise Nonuniformity Correction Framework for Photon-Counting CT in Multi-material Imaging Scenarios

Enze Zhou, Shaoyang Wang, Qian Liu, Wenting Xu, Yuwei Lu, Shangbin Chen, Gang Zheng, Tianwu Xie, Wenjian Li
article en
1 citations

Abstract

Photon-counting computed tomography (PCCT) has demonstrated significant advancements in recent years; however, pixel-wise detector response nonuniformity remains a key challenge, frequently manifesting as ring artifacts in reconstructed images. Existing correction methods exhibit limited generalizability in complex multi-material scenarios, such as contrast-enhanced imaging. This study introduces a Signal-to-Uniformity Error Polynomial Calibration (STEPC) framework to address this issue. STEPC first fits multi-energy projections using a 2D polynomial surface to generate ideal references, then applies a nonlinear multi-energy polynomial model to predict and correct pixel-wise nonuniformity errors. The model is calibrated using homogeneous slab phantoms of different materials, including PMMA, aluminum, and iodinated contrast agents, enabling correction for both non-contrast and contrast-enhanced imaging. Experiments were performed on a custom Micro-PCCT system with phantoms and mouse. Correction performance of STEPC was evaluated using the mean local standard deviation (MLSD) in the projection domain and the ring artifact deviation (RAD) on the reconstructed images. STEPC consistently outperformed existing correction methods in both non-contrast and contrast-enhanced scenarios. It achieved the lowest MLSD and RAD for both phantoms and mouse scans. These results indicate that STEPC provides a robust and practical solution for correcting detector nonuniformity in multi-material PCCT imaging, witch position it as a promising general-purpose calibration framework for photon-counting CT systems.

ArXiv.org
Openalex Percentile: Top 24%
Advanced X-ray and CT Imaging
0.67
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

STEPC: A Pixel-wise Nonuniformity Correction Framework for Photon-Counting CT in Multi-material Imaging Scenarios — Enze Zhou, Shaoyang Wang, et al. · ArXiv.org (2025) | TGRS Research Map | TGRS