Adult brain T1 and T2 values measured at 3 T using magnetic resonance fingerprinting with phantom validation
Phantom measurements characterized the measurement accuracy, repeatability, and reproducibility throughout the study. The MRF T1 and T2 measurements of normal brain tissues at 3 T are validated through concurrent phantom measurements, and mean values were reported for each tissue. This study provides one data point to answer the question of how often phantom quality assessment must be done in tandem with in vivo measurements.