Multispectral optoelectronic sensor to detect peripheral blood pulsatile variations with equivalent performance in light, medium and dark skin tones

Inaccurate oxygen saturation (SpO<sub>2</sub>) measurements using photoplethysmography (PPG) based pulse oximeters on dark skin tones, were a direct cause of increased mortality during the Covid pandemic (2019-2022) and continue to negatively impact routine clinical monitoring. Inaccuracies are commonly interpreted as being caused by rich pigmentation in dark skin tones. A multispectral optoelectronic sensor (mSOS) with bespoke electronics has been created to overcome skin tone related difficulties of capturing peripheral blood pulsatile variations in deoxyhemoglobin (HHb) in red visible spectrum (>600 nm), and to improve the detection of oxyhaemoglobin (HbO<sub>2</sub>) in infrared spectrum (>850 nm). The study demonstrates that a better understanding of the impact of skin tone on complex tissue optics is required to address pulse oximeter performance difficulties arising from skin pigmentation. The study evaluates the relationship between signal quality and skin tones with spectral illuminations at 515 nm, 601 nm, 631 nm and 940 nm. An approved hypoxia protocol with twelve subjects across three different skin tone groups (Individual Typology Angles (ITA): 50<sup>°</sup> - 41<sup>°</sup>, Monk Skin Tones: 2, 3 (Group I), ITA: 23<sup>°</sup> - 1<sup>°</sup>; MST:4, 5 (Group II), and ITA: -31° - -66<sup>°</sup>, MST:8, 9 and 10 (Group III)) was implemented with the mSOS attached to the back of subject's wrist, including six subjects of Group III. The results show the mSOS is capable of detecting peripheral blood perfusion variations with equivalent performance across three skin tone groups for all four wavelengths and throughout the oxygen desaturation range from 100% to 70%. Performance was maintained for all twelve subjects irrespective of skin tone as judged by signal to noise ratio (SNR) with the range of 19.19 - 25.00 and the standard deviation in the range of 1.42 - 5.60. Signal quality (SNR) remains consistently high across all the illumination wavelengths for all subjects, at all desaturation levels (100 - 70%). Upcoming studies will investigate SpO<sub>2</sub> calibration against SaO<sub>2</sub> gold standard reference with a larger subject cohort.

Authors

Institutions

Publication Details

Journal
Scientific Reports
Published
2025-12-13
DOI
https://doi.org/10.1038/s41598-025-31116-9
Primary Topic
Non-Invasive Vital Sign Monitoring
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Multispectral optoelectronic sensor to detect peripheral blood pulsatile variations with equivalent performance in light, medium and dark skin tones

Sijung Hu, V. M. Dwyer, Yasmin Elsahar, Jiajin Hou
Scientific Reports
Non-Invasive Vital Sign Monitoring
article

Multispectral optoelectronic sensor to detect peripheral blood pulsatile variations with equivalent performance in light, medium and dark skin tones

Sijung Hu, V. M. Dwyer, Yasmin Elsahar, Jiajin Hou
article en

Abstract

Inaccurate oxygen saturation (SpO<sub>2</sub>) measurements using photoplethysmography (PPG) based pulse oximeters on dark skin tones, were a direct cause of increased mortality during the Covid pandemic (2019-2022) and continue to negatively impact routine clinical monitoring. Inaccuracies are commonly interpreted as being caused by rich pigmentation in dark skin tones. A multispectral optoelectronic sensor (mSOS) with bespoke electronics has been created to overcome skin tone related difficulties of capturing peripheral blood pulsatile variations in deoxyhemoglobin (HHb) in red visible spectrum (>600 nm), and to improve the detection of oxyhaemoglobin (HbO<sub>2</sub>) in infrared spectrum (>850 nm). The study demonstrates that a better understanding of the impact of skin tone on complex tissue optics is required to address pulse oximeter performance difficulties arising from skin pigmentation. The study evaluates the relationship between signal quality and skin tones with spectral illuminations at 515 nm, 601 nm, 631 nm and 940 nm. An approved hypoxia protocol with twelve subjects across three different skin tone groups (Individual Typology Angles (ITA): 50<sup>°</sup> - 41<sup>°</sup>, Monk Skin Tones: 2, 3 (Group I), ITA: 23<sup>°</sup> - 1<sup>°</sup>; MST:4, 5 (Group II), and ITA: -31° - -66<sup>°</sup>, MST:8, 9 and 10 (Group III)) was implemented with the mSOS attached to the back of subject's wrist, including six subjects of Group III. The results show the mSOS is capable of detecting peripheral blood perfusion variations with equivalent performance across three skin tone groups for all four wavelengths and throughout the oxygen desaturation range from 100% to 70%. Performance was maintained for all twelve subjects irrespective of skin tone as judged by signal to noise ratio (SNR) with the range of 19.19 - 25.00 and the standard deviation in the range of 1.42 - 5.60. Signal quality (SNR) remains consistently high across all the illumination wavelengths for all subjects, at all desaturation levels (100 - 70%). Upcoming studies will investigate SpO<sub>2</sub> calibration against SaO<sub>2</sub> gold standard reference with a larger subject cohort.

Scientific ReportsVol. 16(1)
Loughborough University (GB)
Shenzhen University
Openalex Percentile: Top 61%
Non-Invasive Vital Sign Monitoring
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.