Simulation and experimental characterization of in-beam PETITION PET scanner for proton therapy

An in-beam PET scanner has been developed within the PETITION (PET for InTensive care units and Innovative protON therapy) collaboration to exploit the full potential of proton therapy by enabling either real-time range verification of proton beams or daily biological adaptation of treatment plans based on PET imaging acquired in treatment position before delivery. To optimize the design, the performance of various scanner geometries was first investigated using Monte Carlo simulations in GATE (v8.2), assessing sensitivity and spatial resolution with a 1 MBq 22 Na point source and evaluating image reconstruction using a uniform 1 MBq activity-filled water cylinder and a proton-irradiated head phantom. Based on these studies, a novel C-shaped scanner was developed and simulations were compared with the experimental performance. The prototype achieved a peak simulated sensitivity of 2.9%, with measured sensitivity of 2.1%. Spatial resolution was 2.0 mm FWHM tangentially and axially, and 3.5–4.0 mm radially, showing good agreement between simulations and measurements. Multi-angle acquisition and reconstruction improved image quality, reducing artefacts and enhancing similarity to reference scans in the water cylinder study. Activity distributions from the proton-irradiated head phantom showed alignment with expected activation profiles, demonstrating reliable reconstruction of clinically relevant signals. These findings confirm the feasibility of an open-ring PET scanner for proton therapy with performance comparable to clinical PET systems while uniquely enabling in-treatment position imaging and online beam range verification. Therefore, it supports novel approaches for adaptive and biologically guided proton therapy.

Authors

Institutions

Publication Details

Journal
Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment
Published
2025-12-13
DOI
https://doi.org/10.1016/j.nima.2025.171230
Primary Topic
Radiation Therapy and Dosimetry
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Simulation and experimental characterization of in-beam PETITION PET scanner for proton therapy

Carla Winterhalter, Shubhangi Makkar, Christian Ritzer, Keegan McNamara et al.
Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment
Radiation Therapy and Dosimetry
article

Simulation and experimental characterization of in-beam PETITION PET scanner for proton therapy

Carla Winterhalter, Shubhangi Makkar, Christian Ritzer, Keegan McNamara, Judith Flock, Antony Lomax, G. Dissertori, Marina Béguin, Cristián Fuentes, Noemi Bührer
article en

Abstract

An in-beam PET scanner has been developed within the PETITION (PET for InTensive care units and Innovative protON therapy) collaboration to exploit the full potential of proton therapy by enabling either real-time range verification of proton beams or daily biological adaptation of treatment plans based on PET imaging acquired in treatment position before delivery. To optimize the design, the performance of various scanner geometries was first investigated using Monte Carlo simulations in GATE (v8.2), assessing sensitivity and spatial resolution with a 1 MBq 22 Na point source and evaluating image reconstruction using a uniform 1 MBq activity-filled water cylinder and a proton-irradiated head phantom. Based on these studies, a novel C-shaped scanner was developed and simulations were compared with the experimental performance. The prototype achieved a peak simulated sensitivity of 2.9%, with measured sensitivity of 2.1%. Spatial resolution was 2.0 mm FWHM tangentially and axially, and 3.5–4.0 mm radially, showing good agreement between simulations and measurements. Multi-angle acquisition and reconstruction improved image quality, reducing artefacts and enhancing similarity to reference scans in the water cylinder study. Activity distributions from the proton-irradiated head phantom showed alignment with expected activation profiles, demonstrating reliable reconstruction of clinically relevant signals. These findings confirm the feasibility of an open-ring PET scanner for proton therapy with performance comparable to clinical PET systems while uniquely enabling in-treatment position imaging and online beam range verification. Therefore, it supports novel approaches for adaptive and biologically guided proton therapy.

Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated EquipmentVol. 1084
Paul Scherrer Institute (CH), ETH Zurich (CH)
Openalex Percentile: Top 41%
Radiation Therapy and Dosimetry
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.