A novel wearable dosimeter system that can analyze the incident direction of X-rays for medical dosimetry – Improvements to the detector arrangements and analysis algorithm –

When performing real-time dosimetry using an active-type dosimeter during clinical fluoroscopic procedures, angular dependence of dosimeter response should be taken into account. Our research group addressed this issue and proposed a triple-type dosimeter that can determine the incident angle of scattered X-rays. The triple-type detector consists of three active dosimeters. The two side dosimeters have slope filters to enhance the angular dependence and are intentionally tilted. The central dosimeter faces forward. The incident angle of X-rays ( θ in ) is estimated using the signal differences between the central dosimeter and the left and/or right dosimeters. Then, the absolute dose is determined by correcting the angular dependence of the central dosimeter based on the estimated θ in . In order to verify the concept of the triple-type dosimeter, we conducted a proof-of-concept experiment using clinical X-ray fluoroscopic equipment. Scattered X-rays were generated by irradiating an elliptical cylindrical water phantom. The response of the triple-type dosimeter was evaluated by rotating it to vary the incident angle of scattered X-rays generated by the water phantom. The proposed dosimetry system could estimate the θ in over an angular range of ±80° (with uncertainty of 1.35°), which is 30° wider than the previous version, and successfully determined the absolute doses after correction for the angular dependence of the dosimeter. Although the active-type dosimeter had a systematic uncertainty related to the angular dependence of ±15.2 %, our system succeeded in reducing the systematic uncertainty to ±3.2 %. • Aive-type dosimeters have relatively larger angular dependence. • The triple-type dosimeter can analyze the incident angle and absolute X-ray dose. • The incident angle can be analyzed with an accuracy of 1.35°. • The angle of the incident X-ray can be calculated when the view is within ±80°. • The accuracy of the absolute dose was evaluated to be 3.2 %.

Authors

Publication Details

Journal
Radiation Measurements
Published
2025-12-13
DOI
https://doi.org/10.1016/j.radmeas.2025.107592
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A novel wearable dosimeter system that can analyze the incident direction of X-rays for medical dosimetry – Improvements to the detector arrangements and analysis algorithm –

R Ishii, Hiroaki Hayashi, Natsumi Kimoto, Rina Nishigami et al.
Radiation Measurements
Radiation Dose and Imaging
article

A novel wearable dosimeter system that can analyze the incident direction of X-rays for medical dosimetry – Improvements to the detector arrangements and analysis algorithm –

R Ishii, Hiroaki Hayashi, Natsumi Kimoto, Rina Nishigami, Toshihiro Iguchi, Daiki Kobayashi, Masayuki Honda, Takashi Asahara, Takegami Kazuki, M. Mitani, Sota GOTO
article en

Abstract

When performing real-time dosimetry using an active-type dosimeter during clinical fluoroscopic procedures, angular dependence of dosimeter response should be taken into account. Our research group addressed this issue and proposed a triple-type dosimeter that can determine the incident angle of scattered X-rays. The triple-type detector consists of three active dosimeters. The two side dosimeters have slope filters to enhance the angular dependence and are intentionally tilted. The central dosimeter faces forward. The incident angle of X-rays ( θ in ) is estimated using the signal differences between the central dosimeter and the left and/or right dosimeters. Then, the absolute dose is determined by correcting the angular dependence of the central dosimeter based on the estimated θ in . In order to verify the concept of the triple-type dosimeter, we conducted a proof-of-concept experiment using clinical X-ray fluoroscopic equipment. Scattered X-rays were generated by irradiating an elliptical cylindrical water phantom. The response of the triple-type dosimeter was evaluated by rotating it to vary the incident angle of scattered X-rays generated by the water phantom. The proposed dosimetry system could estimate the θ in over an angular range of ±80° (with uncertainty of 1.35°), which is 30° wider than the previous version, and successfully determined the absolute doses after correction for the angular dependence of the dosimeter. Although the active-type dosimeter had a systematic uncertainty related to the angular dependence of ±15.2 %, our system succeeded in reducing the systematic uncertainty to ±3.2 %. • Aive-type dosimeters have relatively larger angular dependence. • The triple-type dosimeter can analyze the incident angle and absolute X-ray dose. • The incident angle can be analyzed with an accuracy of 1.35°. • The angle of the incident X-ray can be calculated when the view is within ±80°. • The accuracy of the absolute dose was evaluated to be 3.2 %.

Radiation MeasurementsVol. 191
Openalex Percentile: Top 45%
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.