Bench Testing Analysis of Perforation Management in the Setting of Bifurcation Coronary Intervention
In this bifurcation perforation in-bench model, high tip-load wires effectively penetrated covered stents toward excluded branches after crossover covered stent implantation. Conversion to Culotte, using either a DES or a second covered stent, was feasible. Overhanging graft material following SB fenestration could be addressed by overstenting with a DES.
Authors
- Christian Hengstenberg (ORCID: https://orcid.org/0000-0002-8284-2994)
- Judit Andréka (ORCID: https://orcid.org/0009-0009-8068-090X)
- Mohammad Abdelghani (ORCID: https://orcid.org/0000-0001-6949-8749)
- Rayyan Hemetsberger (ORCID: https://orcid.org/0000-0001-5390-0916)
- Johannes Gollmer (ORCID: https://orcid.org/0000-0002-8734-6359)
- A. Mâaroufi (ORCID: https://orcid.org/0000-0001-5662-3766)
- Nader Mankerious (ORCID: https://orcid.org/0000-0003-4809-5371)
- Daniel Lootz
- Kevin Hamzaraj (ORCID: https://orcid.org/0000-0002-2505-8554)
- Tom Johnson (ORCID: https://orcid.org/0000-0003-4638-601X)
- Serdar Farhan (ORCID: https://orcid.org/0000-0002-8237-1280)
- Gábor G. Tóth (ORCID: https://orcid.org/0000-0002-0283-9091)
Institutions
- Al-Azhar University (EG)
- Medical University of Graz (AT)
- Lenox Hill Hospital (US)
- University of Bristol (GB)
- NIHR Bristol Cardiovascular Biomedical Research Unit (GB)
- Segeberger Kliniken (DE)
- Amsterdam University Medical Centers (NL)
- Sohar University (OM)
- Medical University of Vienna (AT)
- Biotronik (Switzerland) (CH)
- University of Amsterdam (NL)
Publication Details
- Journal
- Catheterization and Cardiovascular Interventions
- Published
- 2025-12-12
- DOI
- https://doi.org/10.1002/ccd.70406
- Primary Topic
- Coronary Interventions and Diagnostics
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- Edwards Lifesciences
- Boston Scientific Corporation
- Abiomed
- Abbott Vascular