Clinical performance of the BD Respiratory Viral Panel – SCV2 for BD MAX™ system with nasopharyngeal and anterior nasal specimens

The detection of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) relies heavily on highly sensitive and specific assays. While nasopharyngeal (NP) specimens are considered the gold standard, it is crucial that current assays also support the use of anterior nasal swab (NS) specimens, as they can be more suitable in situations where NP specimen collection is difficult or impractical. Paired NP and NS specimens prospectively collected from symptomatic and asymptomatic participants were utilized to evaluate the BD Respiratory Viral Panel - SCV2 for BD MAX™ System (BD RVP SCV2) clinical performance against a composite comparator of three CE-marked assays: cobas® SARS-CoV-2, Aptima® SARS-CoV-2, and Lyra® SARS-CoV-2. Concordance between at least two assays established a positive or negative comparator result. Each specimen was divided into four aliquots, one for each assay. BD RVP SCV2 performance was assessed by comparing results from NP specimens (NP vs NP), NS specimens (NS vs NS), and NS results with paired NP comparator results (NS vs NP). Positive and negative percent agreements (PPA and NPA) were calculated for all comparisons. Combined symptomatic and asymptomatic specimens, when tested with the BD RVP SCV2 assay, met all set acceptance criteria, with PPAs of 98.7 % (NP vs NP), 98.4 % (NS vs NS), and 92.6 % (NS vs NP) while NPAs ranged from 97.7 % to 98.0 %. These findings confirm that the BD Respiratory Viral Panel - SCV2 for BD MAX System performs well for detecting SARS-CoV-2 in NP and NS specimens from symptomatic and asymptomatic populations.

Authors

Institutions

Publication Details

Journal
Journal of Clinical Virology Plus
Published
2025-12-12
DOI
https://doi.org/10.1016/j.jcvp.2025.100237
Primary Topic
SARS-CoV-2 detection and testing
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Clinical performance of the BD Respiratory Viral Panel – SCV2 for BD MAX™ system with nasopharyngeal and anterior nasal specimens

Jeffry Leitch, Sonia Paradis, Matthew L. Faron, Barbara Van Der Pol et al.
Journal of Clinical Virology Plus
SARS-CoV-2 detection and testing
article

Clinical performance of the BD Respiratory Viral Panel – SCV2 for BD MAX™ system with nasopharyngeal and anterior nasal specimens

Jeffry Leitch, Sonia Paradis, Matthew L. Faron, Barbara Van Der Pol, Karen Yanson, Nathan A. Ledeboer, William Laviers, Elizabeth Lockamy, Thomas E. Davis
article en

Abstract

The detection of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) relies heavily on highly sensitive and specific assays. While nasopharyngeal (NP) specimens are considered the gold standard, it is crucial that current assays also support the use of anterior nasal swab (NS) specimens, as they can be more suitable in situations where NP specimen collection is difficult or impractical. Paired NP and NS specimens prospectively collected from symptomatic and asymptomatic participants were utilized to evaluate the BD Respiratory Viral Panel - SCV2 for BD MAX™ System (BD RVP SCV2) clinical performance against a composite comparator of three CE-marked assays: cobas® SARS-CoV-2, Aptima® SARS-CoV-2, and Lyra® SARS-CoV-2. Concordance between at least two assays established a positive or negative comparator result. Each specimen was divided into four aliquots, one for each assay. BD RVP SCV2 performance was assessed by comparing results from NP specimens (NP vs NP), NS specimens (NS vs NS), and NS results with paired NP comparator results (NS vs NP). Positive and negative percent agreements (PPA and NPA) were calculated for all comparisons. Combined symptomatic and asymptomatic specimens, when tested with the BD RVP SCV2 assay, met all set acceptance criteria, with PPAs of 98.7 % (NP vs NP), 98.4 % (NS vs NS), and 92.6 % (NS vs NP) while NPAs ranged from 97.7 % to 98.0 %. These findings confirm that the BD Respiratory Viral Panel - SCV2 for BD MAX System performs well for detecting SARS-CoV-2 in NP and NS specimens from symptomatic and asymptomatic populations.

Journal of Clinical Virology PlusVol. 6(1)
Becton Dickinson (United States) (US), Medical College of Wisconsin (US), University of Alabama at Birmingham (US), Becton Dickinson (Canada) (CA), Indiana University School of Medicine, Indiana University – Purdue University Indianapolis (US)
Openalex Percentile: Top 54%
SARS-CoV-2 detection and testing
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.