Detection of the Linac Jaw face angle misalignment using high energy electron beam symmetry
We have demonstrated that changes in symmetry of high energy electron beam with 90° collimator rotation can signal that the jaw face angle is not properly aligned with beam divergence.
Authors
- Jared D. Ohrt (ORCID: https://orcid.org/0000-0002-2383-8311)
- Peter Balter (ORCID: https://orcid.org/0000-0003-4806-8920)
- Song Gao (ORCID: https://orcid.org/0000-0003-2743-3466)
- Andres Cibrian
Institutions
- The University of Texas MD Anderson Cancer Center (US)
Publication Details
- Journal
- Journal of Applied Clinical Medical Physics
- Published
- 2025-12-01
- DOI
- https://doi.org/10.1002/acm2.70419
- Primary Topic
- Radiation Effects and Dosimetry
- Type
- article
- Field-Weighted Citation Impact
- 0.00