Compressed sensing based optimization of electromagnetic field measurements required for quality assurance of hyperthermia applicators

<i>Objective</i>. Quality assurance of hyperthermia applicators can be a cumbersome task. Periodic validation of the fields generated by the applicator is crucial for ensuring proper device performance, but the required measurements are very time-consuming. While most clinics use heating rate as a parameter of interest, consensus exists that spatial variation of the electromagnetic field in three dimensions (3D) would be much more insightful. Unfortunately, such 3D coverage would require measurements at many locations.<i>Approach</i>. To address this challenge, we propose a compressed sensing based methodology that enables accurate E-field and specific absorption rate (SAR) reconstruction from significantly reduced sampling densities. Using a Lucite Cone Applicator (LCA) and a homogeneous tissue-mimicking phantom, E-field measurements were obtained via a robotic scanning system equipped with an isotropic EM field probe (EX3DV4, SPEAG). Field maps were reconstructed using a discrete cosine transform (DCT)-based compressed sensing algorithm and evaluated using peak signal-to-noise ratio (PSNR), structural similarity index (SSIM), and the area under 50\%-iso-field contour overlap error. This error refers to the computational reconstruction accuracy of the compressed sensing algorithm when benchmarked against a densely sampled high-resolution reference scan.<i>Main results</i>. Results demonstrated that accurate field reconstruction can be achieved using only 8\% of the full measurements, reducing acquisition time from 135 minutes to just 11 minutes, while maintaining clinically relevant precision (SSIM = 0.9, PSNR = 27 dB, 50\%-iso-field contour overlap error is within ± 2.5\%).<i>Significance</i>. In this way, the need for extensive measurements is reduced while validation reliability is maintained. This approach delivers a faster solution, enhancing information content while significantly reducing the time required for quality assurance in hyperthermia clinics.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2025-12-09
DOI
https://doi.org/10.1088/1361-6560/ae2aa3
Primary Topic
Ultrasound and Hyperthermia Applications
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Compressed sensing based optimization of electromagnetic field measurements required for quality assurance of hyperthermia applicators

Kemal Sümser, Margarethus M. Paulides, Sergio Curto, Deovrat Dilip Phal
Physics in Medicine and Biology
Ultrasound and Hyperthermia Applications
article

Compressed sensing based optimization of electromagnetic field measurements required for quality assurance of hyperthermia applicators

Kemal Sümser, Margarethus M. Paulides, Sergio Curto, Deovrat Dilip Phal
article en

Abstract

<i>Objective</i>. Quality assurance of hyperthermia applicators can be a cumbersome task. Periodic validation of the fields generated by the applicator is crucial for ensuring proper device performance, but the required measurements are very time-consuming. While most clinics use heating rate as a parameter of interest, consensus exists that spatial variation of the electromagnetic field in three dimensions (3D) would be much more insightful. Unfortunately, such 3D coverage would require measurements at many locations.<i>Approach</i>. To address this challenge, we propose a compressed sensing based methodology that enables accurate E-field and specific absorption rate (SAR) reconstruction from significantly reduced sampling densities. Using a Lucite Cone Applicator (LCA) and a homogeneous tissue-mimicking phantom, E-field measurements were obtained via a robotic scanning system equipped with an isotropic EM field probe (EX3DV4, SPEAG). Field maps were reconstructed using a discrete cosine transform (DCT)-based compressed sensing algorithm and evaluated using peak signal-to-noise ratio (PSNR), structural similarity index (SSIM), and the area under 50\%-iso-field contour overlap error. This error refers to the computational reconstruction accuracy of the compressed sensing algorithm when benchmarked against a densely sampled high-resolution reference scan.<i>Main results</i>. Results demonstrated that accurate field reconstruction can be achieved using only 8\% of the full measurements, reducing acquisition time from 135 minutes to just 11 minutes, while maintaining clinically relevant precision (SSIM = 0.9, PSNR = 27 dB, 50\%-iso-field contour overlap error is within ± 2.5\%).<i>Significance</i>. In this way, the need for extensive measurements is reduced while validation reliability is maintained. This approach delivers a faster solution, enhancing information content while significantly reducing the time required for quality assurance in hyperthermia clinics.

Physics in Medicine and BiologyVol. 70(24)
Erasmus MC Cancer Institute (NL), Eindhoven University of Technology (NL), Erasmus University Rotterdam (NL)
Eurostars
Openalex Percentile: Top 61%
Ultrasound and Hyperthermia Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.