Validation and clinical use of the Maquet (Getinge) original series Rotaflow extracorporeal membrane oxygenation device in hyperbaric conditions: a technical report
After safety related modifications to the console's power supply, the Rotaflow 1 console, Quadrox oxygenator and PLS circuit performed satisfactorily up to 243 kPa during repeated clinical use.
Authors
- Vinodh Bhagyalakshmi Nanjayya (ORCID: https://orcid.org/0000-0001-7651-8488)
- Ian Millar
- Theo Tsouras
- Bridget Devaney (ORCID: https://orcid.org/0000-0001-6521-418X)
- Zhiliang Caleb Lin (ORCID: https://orcid.org/0000-0001-9582-7745)
- Lloyd Roberts
- Christopher Covelli
Institutions
- Alfred Health (AU)
- Monash University (AU)
Publication Details
- Journal
- Diving and Hyperbaric Medicine Journal
- Published
- 2025-12-09
- DOI
- https://doi.org/10.28920/dhm55.4.323-329
- Citations
- 1
- Primary Topic
- Mechanical Circulatory Support Devices
- Type
- article
- Field-Weighted Citation Impact
- 0.67