Development of an LC non-resonant detector for mid-infrared photothermal (MIP) microscopy

An inductor-capacitor (LC) non-resonant photodiode detector has been developed and optimized for the signal characteristics of mid-infrared photothermal (MIP) microscopy. The detector incorporates a large photosensitive area (diameter = 5 mm), a flat frequency response, direct current-alternating current (DC-AC) separation, and a single-photodiode configuration. A gain of 2.7 × 104 is achieved for weak AC signals, with a wide-flat -3 dB frequency response ranging from 4.7 kHz to 2.1 MHz. Compared with that of built-in detector of a standard MIP instrument, the signal-to-noise ratio of MIP imaging is increased by ∼5 times. The detector is simple to operate, highly resistant to external noise, and fulfills the imaging requirements of MIP experiments.

Authors

Institutions

Publication Details

Journal
Review of Scientific Instruments
Published
2025-12-01
DOI
https://doi.org/10.1063/5.0300933
Primary Topic
Thermography and Photoacoustic Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Development of an LC non-resonant detector for mid-infrared photothermal (MIP) microscopy

Zhi‐Cong Zeng, Tang-Xiu Li, Ting Pan, Haixia Zhang et al.
Review of Scientific Instruments
Thermography and Photoacoustic Techniques
article

Development of an LC non-resonant detector for mid-infrared photothermal (MIP) microscopy

Zhi‐Cong Zeng, Tang-Xiu Li, Ting Pan, Haixia Zhang, Yuewen Zhang, Jianping Yang, Xiu-Jia Yang, Yan‐Ping Shi
article en

Abstract

An inductor-capacitor (LC) non-resonant photodiode detector has been developed and optimized for the signal characteristics of mid-infrared photothermal (MIP) microscopy. The detector incorporates a large photosensitive area (diameter = 5 mm), a flat frequency response, direct current-alternating current (DC-AC) separation, and a single-photodiode configuration. A gain of 2.7 × 104 is achieved for weak AC signals, with a wide-flat -3 dB frequency response ranging from 4.7 kHz to 2.1 MHz. Compared with that of built-in detector of a standard MIP instrument, the signal-to-noise ratio of MIP imaging is increased by ∼5 times. The detector is simple to operate, highly resistant to external noise, and fulfills the imaging requirements of MIP experiments.

Review of Scientific InstrumentsVol. 96(12)
Lanzhou Institute of Chemical Physics (CN), University of Chinese Academy of Sciences (CN), Lanzhou University (CN)
Openalex Percentile: Top 56%
Thermography and Photoacoustic Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.