S1239 Adverse Events Associated With the CapsoCam Plus (SV-3)™ Capsule Endoscope: A Review of the FDA MAUDE Database

Introduction: CapsoCam Plus (SV-3)™ is a single-use, ingestible capsule endoscope that is used to visualize the small bowel. It uses 4 lateral cameras to capture a 360° view, surpassing other systems like PillCam SB3™, Endocapsule™, MicroCam™, and OMOM HD™ (156°–172°). Unlike other devices, CapsoCam stores images internally, which are retrieved after excretion. It was FDA-cleared in 2019 for remote ingestion under virtual medical supervision. Our study aimed to analyze reported adverse events related to the use of CapsoCam Plus (SV-3)™ using the Food and Drug Administration’s (FDA) Manufacturer and User Facility Device Experience (MAUDE) database. Methods: We analyzed adverse events related to CapsoCam Plus (SV-3)™ reported in the FDA’s MAUDE database from January 2018 to July 2024. Reported claims were categorized into patient-related and device-related issues. A descriptive analysis of these claims was then performed. Results: A total of 18 events were identified: 15 device-related (83.3%) and 3 patient-related (16.7%). The most common device related adverse event was capsule retention (n = 14, 77.7%). Of these, 4 were removed via colonoscopy, 1 with double balloon enteroscopy, 3 with esophagogastroduodenoscopy, and 6 required surgical removal. Two surgical cases followed failed double balloon retrievals. Seven of the retention events (n = 7, 50%) occurred in patients with pre-existing gastrointestinal (GI) conditions: Crohn’s disease (n = 4), ulceration from carcinoid tumor (n = 1), prior mucosal banding (n = 1), and a stricture within a ventral hernia (n = 1, 6.67%). One case of device fragmentation was reported with spontaneous excretion that did not require intervention. All patient-related events (n = 3, 22.3%) involved bowel obstruction: 2 in the small bowel, 1 unspecified. All 3 patients required surgery, including 1 exploratory laparotomy. One patient had been treated with 4 types of laxatives before surgery. Conclusion: CapsoCam Plus (SV-3)™ is a newly approved device for small bowel visualisation. Capsule retention is the most common device related problem, particularly in patients with underlying GI pathology. This highlights that capsule retention continues to be a significant issue across video capsule endoscopy devices. Future efforts should focus on innovative techniques to improve capsule design to reduce retention risk in patients.

Authors

Publication Details

Journal
The American Journal of Gastroenterology
Published
2025-10-01
DOI
https://doi.org/10.14309/01.ajg.0001132416.99734.d1
Primary Topic
Gastrointestinal Bleeding Diagnosis and Treatment
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

S1239 Adverse Events Associated With the CapsoCam Plus (SV-3)™ Capsule Endoscope: A Review of the FDA MAUDE Database

Seyma Bayram, Elmkdad Mohammed, Ahmed Dirweesh, E A Mohammed et al.
The American Journal of Gastroenterology
Gastrointestinal Bleeding Diagnosis and Treatment
article

S1239 Adverse Events Associated With the CapsoCam Plus (SV-3)™ Capsule Endoscope: A Review of the FDA MAUDE Database

Seyma Bayram, Elmkdad Mohammed, Ahmed Dirweesh, E A Mohammed, Mohammad Bilal, Mohammad Adam, Mohamed Abdallah, Sunny Kumar, Khalid Ahmed, Marwa Salman
article en

Abstract

Introduction: CapsoCam Plus (SV-3)™ is a single-use, ingestible capsule endoscope that is used to visualize the small bowel. It uses 4 lateral cameras to capture a 360° view, surpassing other systems like PillCam SB3™, Endocapsule™, MicroCam™, and OMOM HD™ (156°–172°). Unlike other devices, CapsoCam stores images internally, which are retrieved after excretion. It was FDA-cleared in 2019 for remote ingestion under virtual medical supervision. Our study aimed to analyze reported adverse events related to the use of CapsoCam Plus (SV-3)™ using the Food and Drug Administration’s (FDA) Manufacturer and User Facility Device Experience (MAUDE) database. Methods: We analyzed adverse events related to CapsoCam Plus (SV-3)™ reported in the FDA’s MAUDE database from January 2018 to July 2024. Reported claims were categorized into patient-related and device-related issues. A descriptive analysis of these claims was then performed. Results: A total of 18 events were identified: 15 device-related (83.3%) and 3 patient-related (16.7%). The most common device related adverse event was capsule retention (n = 14, 77.7%). Of these, 4 were removed via colonoscopy, 1 with double balloon enteroscopy, 3 with esophagogastroduodenoscopy, and 6 required surgical removal. Two surgical cases followed failed double balloon retrievals. Seven of the retention events (n = 7, 50%) occurred in patients with pre-existing gastrointestinal (GI) conditions: Crohn’s disease (n = 4), ulceration from carcinoid tumor (n = 1), prior mucosal banding (n = 1), and a stricture within a ventral hernia (n = 1, 6.67%). One case of device fragmentation was reported with spontaneous excretion that did not require intervention. All patient-related events (n = 3, 22.3%) involved bowel obstruction: 2 in the small bowel, 1 unspecified. All 3 patients required surgery, including 1 exploratory laparotomy. One patient had been treated with 4 types of laxatives before surgery. Conclusion: CapsoCam Plus (SV-3)™ is a newly approved device for small bowel visualisation. Capsule retention is the most common device related problem, particularly in patients with underlying GI pathology. This highlights that capsule retention continues to be a significant issue across video capsule endoscopy devices. Future efforts should focus on innovative techniques to improve capsule design to reduce retention risk in patients.

The American Journal of GastroenterologyVol. 120(10S2)
Openalex Percentile: Top 45%
Gastrointestinal Bleeding Diagnosis and Treatment
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.