Electrochemical Sensor Based on Polydopamine-Coated Silver Nanowire/Defect-Ridden UiO-66-NH <sub>2</sub> Nanocomposites for Trace Detection of Cd(II) and Pb(II) Ions in Water

Due to growing environmental and health concerns associated with toxic heavy metal ion (HMI) pollution, it becomes necessary to develop fast, highly sensitive, and selective detection methods. Herein, an electrochemical sensor based on a glassy carbon electrode (GCE) modified with a thin film of polydopamine-coated silver nanowire (AgNW@PDA)/defect-ridden UiO-66-NH2 (AgNW@PDA/UiO-66-NH2) nanocomposites was developed for trace detection of cadmium (Cd(II)) and lead (Pb(II)) ions in water. The electrochemical performance of the proposed AgNW@PDAt/UiO-66-NH2/GCE sensor was then optimized by varying the polymerization time (t) of dopamine on the AgNWs. Compared to defect-ridden UiO-66-NH2, AgNW@PDAt/UiO-66-NH2 nanocomposites offered a higher electrochemically active surface area, a better standard heterogeneous electron transfer rate constant, and enhanced sensing performances. The AgNW@PDA24/UiO-66-NH2 nanocomposite (t = 24 h) showed the best electrochemical properties and was further utilized for both individual and simultaneous detection of HMIs using differential pulse voltammetry (DPV). The optimized AgNW@PDA24/UiO-66-NH2/GCE sensor achieved a low limit of detection (LOD) of 25.2 nM for Cd(II) and 9.8 nM for Pb(II), and high sensitivity of 7.18 and 18.2 μA μM–1 cm–2, along with a wide linear range of 0.05–1.1 and 0.05–0.7 μM, respectively, for individual detection. The AgNW@PDA24/UiO-66-NH2/GCE exhibited a higher selectivity for Pb(II) ions in the presence of other interfering metal ions, depicting a recovery rate of ∼97–100% over other HMIs. The sensor further demonstrated good stability, repeatability, and reproducibility. The practical applicability of the developed sensor was also confirmed by the successful detection of Cd(II) and Pb(II) ions in spiked tap water samples. Finally, a paper-based flexible sensor was fabricated using the optimized AgNW@PDA24/UiO-66-NH2 ink, yielding a linear range of 5–60 μM and LOD values of 81 and 153 nM for Pb(II) and Cd(II) ions, respectively, thus confirming its potential for on-site monitoring of HMIs in real water samples.

Authors

Institutions

Publication Details

Journal
ACS Applied Nano Materials
Published
2025-12-09
DOI
https://doi.org/10.1021/acsanm.5c03845
Primary Topic
Electrochemical Analysis and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Electrochemical Sensor Based on Polydopamine-Coated Silver Nanowire/Defect-Ridden UiO-66-NH 2 Nanocomposites for Trace Detection of Cd(II) and Pb(II) Ions in Water

Himanshu Kumar Bhatt, Paramita Das, Fantahun Molla
ACS Applied Nano Materials
Electrochemical Analysis and Applications
article

Electrochemical Sensor Based on Polydopamine-Coated Silver Nanowire/Defect-Ridden UiO-66-NH 2 Nanocomposites for Trace Detection of Cd(II) and Pb(II) Ions in Water

Himanshu Kumar Bhatt, Paramita Das, Fantahun Molla
article en

Abstract

Due to growing environmental and health concerns associated with toxic heavy metal ion (HMI) pollution, it becomes necessary to develop fast, highly sensitive, and selective detection methods. Herein, an electrochemical sensor based on a glassy carbon electrode (GCE) modified with a thin film of polydopamine-coated silver nanowire (AgNW@PDA)/defect-ridden UiO-66-NH2 (AgNW@PDA/UiO-66-NH2) nanocomposites was developed for trace detection of cadmium (Cd(II)) and lead (Pb(II)) ions in water. The electrochemical performance of the proposed AgNW@PDAt/UiO-66-NH2/GCE sensor was then optimized by varying the polymerization time (t) of dopamine on the AgNWs. Compared to defect-ridden UiO-66-NH2, AgNW@PDAt/UiO-66-NH2 nanocomposites offered a higher electrochemically active surface area, a better standard heterogeneous electron transfer rate constant, and enhanced sensing performances. The AgNW@PDA24/UiO-66-NH2 nanocomposite (t = 24 h) showed the best electrochemical properties and was further utilized for both individual and simultaneous detection of HMIs using differential pulse voltammetry (DPV). The optimized AgNW@PDA24/UiO-66-NH2/GCE sensor achieved a low limit of detection (LOD) of 25.2 nM for Cd(II) and 9.8 nM for Pb(II), and high sensitivity of 7.18 and 18.2 μA μM–1 cm–2, along with a wide linear range of 0.05–1.1 and 0.05–0.7 μM, respectively, for individual detection. The AgNW@PDA24/UiO-66-NH2/GCE exhibited a higher selectivity for Pb(II) ions in the presence of other interfering metal ions, depicting a recovery rate of ∼97–100% over other HMIs. The sensor further demonstrated good stability, repeatability, and reproducibility. The practical applicability of the developed sensor was also confirmed by the successful detection of Cd(II) and Pb(II) ions in spiked tap water samples. Finally, a paper-based flexible sensor was fabricated using the optimized AgNW@PDA24/UiO-66-NH2 ink, yielding a linear range of 5–60 μM and LOD values of 81 and 153 nM for Pb(II) and Cd(II) ions, respectively, thus confirming its potential for on-site monitoring of HMIs in real water samples.

ACS Applied Nano MaterialsVol. 8(50)
Indian Institute of Science Education and Research, Bhopal (IN)
Openalex Percentile: Top 60%
Electrochemical Analysis and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.