Variability in the Universality of Electron Ionization Mass Spectrometry Response to Oxygenates in Complex Environmental Mixtures

Electron ionization mass spectrometry (EIMS) has been widely used to measure complex atmospheric and environmental mixtures. The sensitivity of analytes in EIMS is typically obtained through a multipoint calibration curve using commercially available standards, but many analytes of interest do not have pure compounds available, and many samples contain a large fraction of unidentified compounds. To address these limitations in quantification, those analytes are often calibrated using a small number of standards with chemical similarities (e.g., shared functional groups). However, little systematic data are available on the variability of EIMS response factors to environmental analytes. In this work, we compare the EIMS response with a flame ionization detector (FID), a near-universal detector for organics, to quantify the variability in EIMS sensitivity to organics in a complex atmospheric mixture, particularly secondary organic aerosols (SOA). Particle-phase oxidation products from four precursor-oxidant pairs were produced, sampled, and analyzed by gas chromatography using a Thermal Desorption Aerosol Gas Chromatograph (TAG) with simultaneous detection by both FID and EIMS. The signals of FID and EIMS are highly correlated across the chromatogram, and peak areas of individual chromatographic peaks measured by both detectors are also closely correlated. The sensitivity of EIMS varies on average by ∼21% relative to FID for peak-by-peak comparison of oxygenated organic compounds within a complex mixture, suggesting that as long as some information is available about an unknown sample to categorize the analytes in the sample, an EIMS can be treated as a universal detector calibrated using chemically similar analytes.

Authors

Institutions

Publication Details

Journal
Analytical Chemistry
Published
2025-12-08
DOI
https://doi.org/10.1021/acs.analchem.5c04153
Primary Topic
Atmospheric chemistry and aerosols
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Variability in the Universality of Electron Ionization Mass Spectrometry Response to Oxygenates in Complex Environmental Mixtures

B. M. Lerner, Purushottam Kumar, Manjula R. Canagaratna, Chenyang Bi et al.
Analytical Chemistry
Atmospheric chemistry and aerosols
article

Variability in the Universality of Electron Ionization Mass Spectrometry Response to Oxygenates in Complex Environmental Mixtures

B. M. Lerner, Purushottam Kumar, Manjula R. Canagaratna, Chenyang Bi, John T. Jayne, Gabriel A. Isaacman-VanWertz
article en

Abstract

Electron ionization mass spectrometry (EIMS) has been widely used to measure complex atmospheric and environmental mixtures. The sensitivity of analytes in EIMS is typically obtained through a multipoint calibration curve using commercially available standards, but many analytes of interest do not have pure compounds available, and many samples contain a large fraction of unidentified compounds. To address these limitations in quantification, those analytes are often calibrated using a small number of standards with chemical similarities (e.g., shared functional groups). However, little systematic data are available on the variability of EIMS response factors to environmental analytes. In this work, we compare the EIMS response with a flame ionization detector (FID), a near-universal detector for organics, to quantify the variability in EIMS sensitivity to organics in a complex atmospheric mixture, particularly secondary organic aerosols (SOA). Particle-phase oxidation products from four precursor-oxidant pairs were produced, sampled, and analyzed by gas chromatography using a Thermal Desorption Aerosol Gas Chromatograph (TAG) with simultaneous detection by both FID and EIMS. The signals of FID and EIMS are highly correlated across the chromatogram, and peak areas of individual chromatographic peaks measured by both detectors are also closely correlated. The sensitivity of EIMS varies on average by ∼21% relative to FID for peak-by-peak comparison of oxygenated organic compounds within a complex mixture, suggesting that as long as some information is available about an unknown sample to categorize the analytes in the sample, an EIMS can be treated as a universal detector calibrated using chemically similar analytes.

Analytical ChemistryVol. 97(50)
Aerodyne Research (US), Virginia Tech (US)
Openalex Percentile: Top 57%
Atmospheric chemistry and aerosols
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.