Development and evaluation of a mechanical chest compression device for standardized rodent cardiopulmonary resuscitation

Small animal models are indispensable in cardiopulmonary resuscitation (CPR) research. High-quality CPR, characterized by consistent chest compression rate, depth, and positioning is crucial for survival. However, achieving standardization in manual high-frequency chest compressions in small animal models remains technically challenging. This study evaluated the reproducibility of manual chest compressions and introduced a novel mechanical chest compression device (MCD) designed to improve consistency in rodent experiments. In an in vitro setup, manual compressions were performed by ten participants at target rates ranging from 100 to 260 bpm, guided by a metronome. Compressions performed on a fluid-filled polymer reservoir were analyzed for the compression rate, variability, and time within a ± 10% target range. A color indicator was used to assess the variability of the compression point. A small animal MCD was designed and tested under the same conditions. In vivo, 5 Sprague-Dawley rats underwent 5 min of electrically induced normothermic cardiac arrest followed by 8 min of external chest compressions using the MCD. Obtained data was compared to the in vitro results. A total of 21,650 manual and 20,098 mechanical compressions were analyzed. At 200 bpm, chest compressions using the MCD were significantly more precise (201 ± 1.2 bpm) than manual compressions (218 ± 21 bpm, p < 0.001) with a significant reduced compression point variability (1.7 ± 0.1 cm<sup>2</sup> vs. 10.8 ± 3.1 cm<sup>2</sup>, p < 0.001). Manual compressions maintained target rate in 58.8% of time compared to 100% for the MCD. In vivo testing confirmed these findings with chest compressions remaining within the target range 100% of the time and showing minimal rate variability (1.8 ± 1.7 bpm). These results highlight the limitations of manual chest compressions and demonstrate the potential of the MCD to enhance standardization and reproducibility in rodent CPR research.

Authors

Institutions

Publication Details

Journal
Scientific Reports
Published
2025-12-08
DOI
https://doi.org/10.1038/s41598-025-31959-2
Primary Topic
Cardiac Arrest and Resuscitation
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Development and evaluation of a mechanical chest compression device for standardized rodent cardiopulmonary resuscitation

Felix Patricius Hans, Sonja Bröer, Sam Brixius, Jakob Wollborn et al.
Scientific Reports
Cardiac Arrest and Resuscitation
article

Development and evaluation of a mechanical chest compression device for standardized rodent cardiopulmonary resuscitation

Felix Patricius Hans, Sonja Bröer, Sam Brixius, Jakob Wollborn, Martin Czerny, Georg Trummer, Hans-Jörg Busch, Jan‐Steffen Pooth, Johannes Dinkelaker, Rita Sanchez-Brandelik, Katharina Denzer
article en

Abstract

Small animal models are indispensable in cardiopulmonary resuscitation (CPR) research. High-quality CPR, characterized by consistent chest compression rate, depth, and positioning is crucial for survival. However, achieving standardization in manual high-frequency chest compressions in small animal models remains technically challenging. This study evaluated the reproducibility of manual chest compressions and introduced a novel mechanical chest compression device (MCD) designed to improve consistency in rodent experiments. In an in vitro setup, manual compressions were performed by ten participants at target rates ranging from 100 to 260 bpm, guided by a metronome. Compressions performed on a fluid-filled polymer reservoir were analyzed for the compression rate, variability, and time within a ± 10% target range. A color indicator was used to assess the variability of the compression point. A small animal MCD was designed and tested under the same conditions. In vivo, 5 Sprague-Dawley rats underwent 5 min of electrically induced normothermic cardiac arrest followed by 8 min of external chest compressions using the MCD. Obtained data was compared to the in vitro results. A total of 21,650 manual and 20,098 mechanical compressions were analyzed. At 200 bpm, chest compressions using the MCD were significantly more precise (201 ± 1.2 bpm) than manual compressions (218 ± 21 bpm, p < 0.001) with a significant reduced compression point variability (1.7 ± 0.1 cm<sup>2</sup> vs. 10.8 ± 3.1 cm<sup>2</sup>, p < 0.001). Manual compressions maintained target rate in 58.8% of time compared to 100% for the MCD. In vivo testing confirmed these findings with chest compressions remaining within the target range 100% of the time and showing minimal rate variability (1.8 ± 1.7 bpm). These results highlight the limitations of manual chest compressions and demonstrate the potential of the MCD to enhance standardization and reproducibility in rodent CPR research.

Scientific ReportsVol. 15(1)
Brigham and Women's Hospital (US), Harvard University (US), University of Freiburg (DE), University Medical Center Freiburg (DE), Universitäts-Herzzentrum Freiburg-Bad Krozingen (DE), Freie Universität Berlin (DE)
American Heart Association, Heart and Stroke Foundation of Canada, Society for Academic Emergency Medicine, European Resuscitation Council
Openalex Percentile: Top 45%
Cardiac Arrest and Resuscitation
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.