Clinical Evaluation of a Novel Point‐of‐Care Hematology Analyzer for Complete Blood Count With Differential
HemoScreen demonstrated generally acceptable agreement with the Sysmex XN for most parameters. However, systematic differences in HCT, MCV, neutrophil, and monocyte results warrant further investigation to assess clinical impact before broader implementation.
Authors
- Ryan C. Shean (ORCID: https://orcid.org/0000-0001-7855-4976)
- Sterling T. Bennett (ORCID: https://orcid.org/0000-0002-5348-6185)
Institutions
- University of Utah (US)
- Intermountain Medical Center (US)
- ARUP Laboratories (United States) (US)
Publication Details
- Journal
- International Journal of Laboratory Hematology
- Published
- 2025-12-08
- DOI
- https://doi.org/10.1111/ijlh.70032
- Primary Topic
- Clinical Laboratory Practices and Quality Control
- Type
- article
- Field-Weighted Citation Impact
- 0.00