Verification, validation, and uncertainty quantification of finite element analysis results for pedicle screw assemblies under ASTM F1717 flexion and extension testing
These findings highlight the importance of accurately defining contact conditions and experimentally characterizing interface behavior. To ensure predictive accuracy and regulatory relevance, validated friction models should be applied and contact parameters precisely calibrated, which can enhance the credibility of spinal biomechanics simulations. Relative to prior F1717-based simulations, this work quantifies the dominant impact of interface modeling and provides actionable parameter bounds for validation (μ = 0.10-0.20 for the tested constructs).